92
Message Reference
(3) Synchronized commands
Set the OPC of the SESR register when finished with all pending operations
Syntax
Command
*
OPC
Description
Sets “OPC bit 0” of the Standard Event Status Register (SESR) when the present
operation is completed.
The next command does not execute until the present operation is completed.
Example
*
OPC
*
ESR?
1
Wait for the present operation to complete and respond with 1
Query
*
OPC?
Response
1
Description
Waits for the present operation to complete and responds with 1.
The next command does not execute until the present operation is completed.
Example
:ROUT:OPEN
*
OPC?
1
Responds with 1 when all relays are opened and all commands are completed.
Wait for the present operation to complete
Syntax
Command
*
WAI
Description
Waits for the present operation to complete.
The next command does not execute until the present operation is completed, however,
the
:ABORt
and
*
TRG
commands are accepted.
Unlike
*
OPC
, the Standard Event Status Register (SESR) is not set. Everything else is the
same.
(4) Status and event control commands
Clear the Event Registers and the Status Byte Register
Syntax
Command
*
CLS
Description
Clears the Event Status Registers. The Status Byte Register bits corresponding to the
Event Status Registers are also cleared. Errors are also cleared.
Set and query Standard Event Status Enable Register (SESER)
Syntax
Command
*
ESE <0
–
255(NR1)>
Query
*
ESE?
Response
<0 to 255 (NR1)>
Description
The SESER mask is set to a numerical value from 0 to 255. The initial value when the
power is turned on is 0. When an event applicable to the specified bit occurs, bit 5, the
Event Status Bit (ESB) of the Status Byte Register (STB) is set.
bit 7
bit 6
bit 5
bit 4
bit 3
bit 2
bit 1
bit 0
PON
URQ
CME
EXE
DDE
QYE
RQC
OPC
Example
*
ESE 36
Sets bit 5 and 2 of the SESER.
Содержание SW1001
Страница 2: ......
Страница 42: ...38 Precautions for Measurement...
Страница 48: ...44 Resetting Scan Operation...
Страница 50: ...46 Checking the Device Status...
Страница 52: ...48 Initialization Settings...
Страница 118: ...114 Sample Programs 3 Select File Save All 4 Check the save location and click Save as...
Страница 156: ...152 Outline Drawings 12 4 Outline Drawings SW1001 Unit mm Tolerance 0 2...
Страница 157: ...153 Outline Drawings SW1002 Unit mm Tolerance 0 2 12 Appendix...
Страница 158: ...154 Outline Drawings...
Страница 161: ...16 01 EN...
Страница 162: ......
Страница 163: ......
Страница 164: ......