127
List of Device Error Numbers
10.2 List of Device Error Numbers
This section provides a list of device error numbers.
You can acquire the number of any error that occurs as well as a description of the error using the
:SYSTem:ERRor?
query.
See: "Query and clear error information" (p. 100)
Error No. Error description
Details
Solution
Reference
−100
Command error
Illegal command string
Check whether there is no error
in the command string.
−200
Execution error
Unable to execute
command with the
specified conditions
It may not be possible to
execute the command in
question under the present
conditions.
−220
Parameter error
Out-of-range or illegal
parameter
Check the command parameter
range or string.
p. 103
−222
Bad Slot/Ch
Illegal slot or channel
specification
Check the specification method
and parameter string.
−313
Backup lost
Corrupt device hardware
information
Repair is necessary.
–
−315
Setting backup lost
Corrupt settings backup
data
−330
Self-test failed
Error during self-test
−360
Communication error Communications error
Check the communications
settings (forward speed,
etc.). Check whether the
communications lines are being
affected by noise.
−361
Rs232c parity error
RS-232C parity error
−362
Rs232c framing error RS-232C framing error
−363
Rs232c overrun
RS-232C overrun error
−365
Time out error
Timeout
−370
Module backup
failed
Corrupt module information Repair is necessary.
–
−371
Comm transfer
Timeout
Communications
forwarding function timeout
Check the communications
settings on the connected
instrument. Check the
command being sent for errors.
−372
Comm transfer
overrun
Communications
forwarding function receive
buffer full
−373
Comm Transfer
RS232C error
Communications
forwarding function
RS-232C communications
error
−400
Query error
Full send buffer when
sending a query
Be sure to receive the response
when you send a query.
10
Maintenance and Service
Содержание SW1001
Страница 2: ......
Страница 42: ...38 Precautions for Measurement...
Страница 48: ...44 Resetting Scan Operation...
Страница 50: ...46 Checking the Device Status...
Страница 52: ...48 Initialization Settings...
Страница 118: ...114 Sample Programs 3 Select File Save All 4 Check the save location and click Save as...
Страница 156: ...152 Outline Drawings 12 4 Outline Drawings SW1001 Unit mm Tolerance 0 2...
Страница 157: ...153 Outline Drawings SW1002 Unit mm Tolerance 0 2 12 Appendix...
Страница 158: ...154 Outline Drawings...
Страница 161: ...16 01 EN...
Страница 162: ......
Страница 163: ......
Страница 164: ......