Command Message List
11
3. Others
Command
Description
Formats
RBT?
Query the battery low status
d1 (Query of battery low status: 0 to 1)
0: Battery low status not detected
1: Battery low status detected
[Format]
RBT?
[Response]
d1: NR1 format
RRJ?
Query the temperature probe connection status
This command is valid only in TC:RJ sourcing or tempera-
ture measurement.
d1 (Query of temperature probe status: 0 to 1)
0: Temperature probe not connected, the input tempera-
ture is outside the measurement range (of -25°C to
80°C).
1: Temperature probe connected
[Format]
RRJ?
[Response]
d1: NR1 format
ERR?
Query the error register.
d1 (error register query: 0 to 255)
bit7: unused
bit6: MLE
Message length error
bit5: HDE
Message header error
bit4: DFE
Data notation error
bit3: DRE
Data range error
bit2: CNE
Command not enforceable
bit1: ISE
Internal signal error
bit0: BDE
Environment data corrupted
[Format]
ERR?
[Response]
d1: NR1 format
*IDN?
Query the Device ID.
d1: Query the Device ID.
Manufacturer's name, Model name, Software version
(Response example)
HIOKI,SS7012, Ver 1.01
[Format]
*IDN?
[Response]
d1: ASCII format
Do not turn off the power switch of SS7012 while executing MEM, MRM or SCN
command. Do not execute these commands while Battery low warning indicator
flashing. These may damage SS7012 and cause the internal error.
Содержание SS7012
Страница 1: ...DC SIGNAL SOURCE Remote Operation Manual SS7012 March 2010 Revised edition 1 SS7012A983 01 10 03H ...
Страница 2: ......
Страница 4: ...Contents ii ...
Страница 17: ...MEMO ...
Страница 18: ......
Страница 19: ......
Страница 20: ......