19
Connecting the Ground Terminal
Example equivalent circuit diagram for ground common
SP7100 CAN Interface rear
AC coupling
The Non-Contact CAN Sensor and the CAN Interface are AC-coupled.
This design serves to prevent a short-circuit in the event the Non-Contact CAN Sensor’s sensor
ground or a Signal Probe’s guard hook comes into contact with a DC power supply (such as a
battery).
Note that there is no AC coupling between the Signal Probe and the sensor ground, as illustrated in
the equivalent circuit diagram, and exercise caution to avoid creating a short-circuit.
Short-circuit protection provided by AC coupling functions only for DC voltages.
HIOKI SP7001A961-01