TEST FIXTURE

Instruction Manual

June 2022  Edition 1
SA9002A960-00  22-06H

SA9002

EN

www.hioki.com/

Warranty

Warranty malfunctions occurring under conditions of normal 

use in conformity with the Instruction Manual and Product 

Precautionary Markings will be repaired free of charge. This 

warranty is valid for a period of three (3) years from the date 

of purchase. Please contact the distributor from which you 

purchased the product for further information on warranty 

provisions.

Introduction

Thank you for choosing the Hioki SA9002 Test Fixture. To 

ensure your ability to get the most out of this instrument over 

the long term, please read this manual carefully and keep it 

available for future reference.

Downloading Slurry Analytical System 

Instruction Manual

This document is the instruction manual of the 

SA9002 Test Fixture.

Make sure to visit our website to download the 

Instruction Manual of the Slurry Analytical System 

and familiarize yourself with it.

https://www.hioki.com/global/products/specialized-solutions/

advanced-tech/id_1078935#downloads

Overview

The SA9002 is a test fixture for the Slurry Analytical System 

with which the SA9001 Electrode Cell can be mounted.

The Test Fixture can be installed to the IM3536 LCR Meter, 

IM3536-01 LCR Meter, and IM3570 Impedance Analyzer.

See the Slurry Analytical System Instruction Manual for detail 

information including the precautions for use of the system.

Symbols and Abbreviations

Safety

This manual classifies seriousness of risks and hazard levels 

as described below.

 

DANGER

Indicates an imminently hazardous 

situation that, if not avoided, will result in 

death or serious injury.

 WARNING

Indicates a potentially hazardous situation 

that, if not avoided, could result in death 

or serious injury.

 CAUTION

Indicates a potentially hazardous situation 

that, if not avoided, could result in minor 

or moderate injury or potential risks of 

damage to the supported product (or to 

other property).

Indicates a prohibited action.

Indicates a mandatory action.

Symbols for standards

Indicates that the product complies with standards 

imposed by EU directives.
Indicates that the product is subject to the Waste 

Electrical and Electronic Equipment (WEEE) Directive 

in EU member nations. Dispose of the product in 

accordance with local regulations.

Checking Package Contents

 SA9002 Test Fixture .........................................................1

 Compensation shorting plate............................................1

 Instruction Manual (this document) ..................................1

Precautions for Use

 DANGER

 

„

Do not use the device to measure samples that 

exceed its ratings or specifications.

Doing so could cause damage to the device or 

overheating, resulting in serious bodily injury.

 

„

Inspect the device and verify proper operation 

before use.

Use of the device while malfunctioning could result in 

serious bodily injury. If you find any damage, contact 

your authorized Hioki distributor or reseller.

 WARNING

Do not install the device in locations such as the 
following:

 

„

In locations where it would be subject to direct 

sunlight or high temperatures

 

„

In locations where it would be exposed to 

corrosive or explosive gases

 

„

In locations where it would be exposed to 

powerful electromagnetic radiation or close to 

objects carrying an electric charge

 

„

Close to inductive heating devices (high-

frequency inductive heating devices, IH cooktops, 

etc.)

 

„

In locations characterized by a large amount of 

mechanical vibration

 

„

In locations where it would be exposed to water, 

oil, chemicals, or solvents

 

„

In locations where it would be exposed to high 

humidity or condensation

 

„

In locations with an excessive amount of dust

Doing so could damage the device or cause it to  

malfunction, resulting in bodily injury.

Cleaning

 CAUTION

 

„

If the device becomes dirty, wipe the device clean 

with a soft cloth moistened with water or a neutral 

detergent.

Never use solvents such as benzene, alcohol, 

acetone, ether, ketone, thinners or gasoline. Doing so 

could deform and discolor the instrument.

When the electrode of the device is stained, gently wipe it 

using a soft cloth with a small amount of alcohol.

Specifications

Measurable 

frequency

DC to 10 MHz

Connectible 

sample

SA9001 Electrode Cell

Residual 

impedance

Residual resistance during short circuit

200 mΩ or less (reference for 100 Hz)

Inter-electrode stray capacitance
0.2 pF or less (reference for 1 MHz)

Dimensions

Approx. 98W × 38H × 24D mm (3.86″W × 1.50″D × 
0.94″H, excluding protruding parts)

Weight

Approx. 210 g (7.4 oz.)

Connection method

BNC connector

Operating 

temperature and 

humidity range

23°C ±5°C (73°F ±9°F)
80% RH or less (non-condensing)

Storage 

temperature and 

humidity range

−10°C to 50°C (−14°F to 122°F)

 80% RH or less (non-condensing)

Operating 

environment

Indoors, altitude up to 2000 m (6562 ft.)

Accessories

Instruction Manual, shorting plate for compensation

Product warranty 

duration

3 years

Parts Names

1

2

4

3

No.

Name

Function

1

SA9001 mounting part

Used to mount the SA9001 Electrode 
Cell.

2

Compensation shorting 
plate

Used to perform short compensation.

3

Lever

Used to fix the SA9002 Test Fixture to 

the IM3536.

4

Connection terminals

Used to connect the Test Fixture to 
the measurement terminals of the 
measuring instrument.

How To Install the Test Fixture

Check that the main 

power switch (rear) of the 

measuring instrument is 

set to off (○).

Face up the SA9001 

mounting part of the 

device, and insert 

the device into the 

measurement terminals of 

the measuring instrument.

Turn the levers on the 

left and right sides of the 

device clockwise to fix the 

device to the measuring 

instrument.

Open Compensation, 

Short Compensation

Performing open compensation and short compensation is 

recommended after the measuring instrument is turned on.

See “3.3 Correcting Errors (Open Compensation, Short 

Compensation)” in the Slurry Analytical System Instruction 

Manual for more details.

Open compensation

Use the SA9001.

Short compensation

Use the shorting plate for 

compensation

SA9001  

Electrode Cell

Shorting plate for 

compensation

HIOKI SA9002A960-00

Содержание SA9002

Страница 1: ...たり発熱したりし 重大な人身事故を 引き起こすおそれがあります 使用前に本器を点検し 本器が正常に動作することを 確認する 本器が故障したまま使用すると 重大な人身事故を引き 起こすおそれがあります 故障を確認した場合は お買 上店 代理店 か最寄りの営業拠点にご連絡ください 警 告 次のような場所に本器を設置しない 直射日光があたる場所 高温になる場所 腐食性ガスや爆発性ガスが発生する場所 強力な電磁波を発生する場所 帯電しているものの近く 誘導加熱装置 高周波誘導加熱装置 IH調理器具など の近く 機械的振動が多い場所 水 油 薬品 溶剤などがかかる場所 多湿 結露する場所 ほこりが多い場所 本器が破損したり誤動作をしたりし 人身事故を引 き起こすおそれがあります クリーニング 注 意 本器の汚れをとるときは 柔らかい布に水か中性洗剤 を少量含ませて 軽く拭く ベンジン アルコール ア...

Страница 2: ...l the device in locations such as the following In locations where it would be subject to direct sunlight or high temperatures In locations where it would be exposed to corrosive or explosive gases In locations where it would be exposed to powerful electromagnetic radiation or close to objects carrying an electric charge Close to inductive heating devices high frequency inductive heating devices I...