Introduction

Thank you for choosing the Hioki RM9010-01, RM9010-02 Four-Point Array 

Probe.To ensure your ability to get the most out of this product over the 

long term, please read this manualcarefully and keep it available for future 

reference.

Overview

The RM9010-01/RM9010-02 Four-Point Array Probe is the probe assembly 

that can plug into the RM3545, RM3545-01, and RM3545-02 Resistance 

Meter to perform the four-point probe resistivity measurements. You can 

employ the four-point probe method to measure resistivity and conductivity 

using the PC application designed for the RM3545 exclusively. The 

application can be installed using the installer contained in the CD that came 

with the instrument.

Initial Inspection

When you open the package, carefully inspect the product to ensure 

that everything is in good condition, and that no damage occurred during 

shipping.  If the product seems to have been damaged or does not work as 

specified, contact your authorized Hioki distributor or reseller.

Maintenance and Service

 •To clean the product, wipe it gently with a soft cloth moistened with water or 

mild detergent. Never use solvents such as benzene, alcohol, acetone, ether, 

ketones, thinners or gasoline, as they can deform and discolor the case.

 •

If the product seems to be malfunctioning, contact your authorized Hioki 

distributor or reseller.

Observe the following when shipping the product.

 •Use the packaging in which the product was initially delivered and then 

pack that in an additional box. Accidental damage suffered in transit is not 

covered by the warranty.

 •Attach a description of the malfunction when sending the product for repair.

Safety notations

In this manual, the   symbol indicates particularly important 
information that the user should read before using the product.

In this document, the severity levels of risk and hazard are classified as 

follows.

 

DANGER

Indicates an imminently hazardous situation that, if not 

avoided, will result in death or serious injury.

 

CAUTION

Indicates a potentially hazardous situation that, if not 

avoided, could result in minor or moderate injury.

Safety Information

Carefully read the instruction manual of Resistance Meter as well as the 

following safety notes before using the products.

 DANGER

Mishandling the product could result in bodily injury or 

even death, as well as damage to the product. Familiarize 

yourself with the instructions and precautions in this 

manual before use.

Precautions for Use

Observe the following precautionary information to ensure that the product 

can be used safely and in a manner that allows it to perform as described in 

its specifications.

 DANGER

 

To prevent an electric shock, be careful to avoid shorting 

live lines with the probe tip.

 

Do not use the product to measure circuits that exceed 

its ratings or specifications. Damage to the product or 

overheating can cause bodily injury.

 

Do not use the product where it may be exposed to oil, 

chemicals, or solvents. Contact with these substances 

may cause cracking in the product, resulting in damage 

or an electric shock.

 

To prevent an electric shock and short-circuit, turn off all 

the instruments before connecting probes.

 CAUTION

 •

To prevent damage due to snapped wires, do not bend or pull 

the base of the probes.

 CAUTION

 • To avoid damage to the product, do not subject it to vibration or 

mechanical shock during transportation and handling. Exercise 

particular care to avoid subjecting the product to mechanical shock, 

for example, by dropping.

 • Do not curl the cable near the cable lock. It may damage the core 

wire or the covering.

 • Do not bend it continuously.

Part Names

RM9010-01

Conductive contact pins 

(5.0 mm pitches)

Cable

Guard plug

Red

Black

SENSE connectors

SOURCE connectors

Protective cover

RM9010-02

Conductive contact pins 

(1.5 mm pitches)

Cable

Cable lock

Protective covers are placed over the pins for protection 

during transport. Remove the sleeve before use.

SENSE connectors

Red
Black

SOURCE connectors

Guard plug

Protective cover

Enlarged view of pins
RM9010-01

RM9010-02

3.9

φ

0.77

5.0

Unit: mm

2.55

1.5

φ

0.5

Procedure

Preliminary Checks

Check the product for any damage that may have occurred during storage or 

shipping, and perform functional checks before use. If you find any damage 

to the product, please contact your authorized Hioki distributor or reseller for 

repair.

Points to check include the pin operation and whether the pins and cable lock 

are loose. As loose screwing of the cable lock and other components can 

result in damage, be sure to tighten them securely before use.

 DANGER

Confirm that the insulation on cables are undamaged and 

that no bare conductors are improperly exposed before 

use. If you find any damage to the product that leads to 

an electric shock, please contact your authorized Hioki 

distributor or reseller for repair.

Make sure the instrument to which the probe is being connected 

turned off.

Connect the probe to the input terminal of the 

instrument. 

Plug the connectors into the instrument’s 

measurement terminals, observing that the red 

connector’s triangle mark is aligned with the 

instrument’s red triangle mark, and the black 

connector’s one with the black one. Next, insert the 

guard plug into the GUARD terminal.

Place conductive contact pins into contact with an object under 

measurement.

Attach the protective sleeve when not in use.

Specifications

Probe material

Base material 

Beryllium copper

Plating 

Gold (primary nickel plating)

Probe tip shape

Spherical

Probe array

Collinear

Spacing between 

adjacent probes

RM9010-01 

5.0 mm ±0.1 mm

(with every probe thrust against an object)

RM9010-02 

1.5 mm ±0.1 mm

(with every probe thrust against an object)

Insulation 

resistance 

between adjacent 

probes

10 G

Ω

 or more for an applied voltage of 25 V at an 

ambient temperature of 23°C and relative humidity 

of 35% RH (reference value)

Probe spring force  

(per spring)

RM9010-01 

About 1.25 N

(with every probe thrust against an object)

RM9010-02 

About 0.85 N

(with every probe thrust against an object)

Maximum rated 

terminal-to-ground 

voltage

30 V AC rms or less, 42.4 V AC peak or less, 60 V 

DC or less

Rated current

RM9010-01 

3 A AC/DC continuous

RM9010-02 

1.5 A AC/DC continuous

Operating 

temperature and 

humidity range

0°C to 40°C (32°F to 104°F), 80% RH or less

(non-condensing)

Storage 

temperature and 

humidity range

−10°C to 50°C (14°F to 122°F), 80% RH or less

 

(non-condensing)

Operating 

environment

Indoors, pollution degree 2,  

operating altitude: up to 2000 m (6562 ft.)

Dimensions

About 1500 mm (59.06″)

Weight

RM9010-01 

About 200 g (7.1 oz.)

RM9010-02 

About 150 g (5.3 oz.)

Product warranty
period

1 year

Accessories

Instruction Manual

CD (Application software)

Replacing the Tip Pin

You can replace conductive contact pins that have been broken or worn with 

new ones.

To order the conductive contact pins, please contact your authorized Hioki 

distributor or reseller.

You will need: long-nose pliers, tweezers

Tu

rn off the instrument and remove the measurement probe.

Pull out a conductive contact pin you wish to replace with long-

nose pliers.

Insert a new conductive contact pin into the socket and push 

the pin with tweezers until the pin tip is flush with the pin-base 

surface.

RM9010-01

RM9010-02

Pin-base surface

Socket

Pin-base surface

Socket

Check the products for proper operation.

Before putting the products to practical use, measure several objects 

of known resistance to check if the products can yield acceptable 

measured values.

www.hioki.com/

FOUR-POINT ARRAY PROBE

Instruction Manual

June 2022  Revised edition 2
RM9010A960-02  22-06H

RM9010-01

RM9010-02

EN

HIOKI RM9010A960-02