Hioki MR8880 Скачать руководство пользователя страница 7

7

Screen display

Display

5.7-inch VGA-TFT color LCD (640 × 480dot)

Waveform display 

scale

Time axis: × 10 to × 2 (zoom view supported for high-speed 

recording only), × 1, × 1/2 to × 1/2,000

Voltage axis: × 20 to × 2,  × 1,  × 1/2 to × 1/10

Comment input

Titles and comments input for individual channels

Logic waveform dis-

play

Select 2 recording widths; display positions can be set separately

Display items

Waveform display; simultaneous display of waveform and gage; 

simultaneous display of waveform, gage, and settings; simultaneous 

display of waveform and calculation results; simultaneous display of 

waveform and cursor values (A/B cursor values) 

The following display items are supported when using real-time 

functionality:

Monitor function

Value 

(instantaneous value or RMS value)

 and measured waveform 

(monitor screen display with refresh rate of 0.5 sec)

Display digits: 5

Instantaneous value 

display

Time: Display of time elapsed since start of measurement or trigger point

Date: Display of date and time at which data was captured

Number of data points: Display of number of data points captured since 

start of measurement

Other display 

functions

• Cursor measurement 

(two cursors [A/B], support for all channels)

• Upper and lower limits can be set 

(to align waveform amplitude with 

upper and lower limits).

• The zero position of the analog waveform can be moved in 1% steps.

• The waveform display can be set to any of 24 colors.

• Zero adjustment can be performed for all channels and ranges at once.

Specifications of Options

 (sold separately)

Appearance and Dimensions

with PRINTER UNIT MR9000 attached

303 mm

(11.93 in)

19

9 m

m

(7

.8

3  in

)

205 mm

(8.07 in)

19

9 m

m

(7

.8

3  in

)

67 m

m

(2

.6

4 i

n)

Analog input 

(Isolated BNC terminal)

CF card slot

Logic probe 

terminal

USB terminal

(for PC communication)

USB terminal

(for USB memory)

External 

control 

terminal

n

  PC Software Specifications   

Bundled with the MR8880 in the CD-R

Wave Viewer (Wv) Software 

Functions

• Simple display of waveform file 

• Text conversion: convert binary data file to text format, with 

selectable space or tab separators in addition to CSV, and specifiable 

section, thinning available 

• Display format settings: scroll functions, enlarge/reduce display, 

display channel settings 

• Others: voltage value trace function, jump to cursor/trigger position 

function

Operating environment

Windows 8/7 (32/64-bit), Vista (32-bit), XP, 2000

Cable length and mass:

 Main unit cable 1.5 m (4.92 ft), input section cable 30 cm 

(0.98 ft), approx. 150 g (5.3 oz) 

Note: The unit-side plug of the 

9320-01

 is different from the 

9320

.

LOGIC PROBE 9320

-01

Function

Detection of voltage signal or relay contact signal for High/Low state recording

Input

4 channels 

(common ground between unit and channels)

, digital/contact input, 

switchable 

(contact input can detect open-collector signals)

Input resistance: 1 MΩ 

(with digital input, 0 to +5 V)

                             500 kΩ or more 

(with digital input, +5 to +50V)

Pull-up resistance: 2 kΩ 

(contact input: internally pulled up to +5 V)

Digital input threshold

1.4V/ 2.5V/ 4.0V

Contact input 

detection resistance

1.4 V: 1.5 kΩ or higher 

(open)

 and 500 Ω or lower 

(short)

2.5 V: 3.5 kΩ or higher 

(open)

 and 1.5 kΩ or lower 

(short)

4.0 V: 25 kΩ or higher 

(open)

 and 8 kΩ or lower 

(short)

Response speed

500ns or lower

Max. allowable input

0 to +50V DC 

(the maximum voltage that can be applied across input pins without 

damage)

LOGIC PROBE MR9321

-01

Function

Detection of AC or DC relay drive signal for High/Low state recording

Can also be used for power line interruption detection

Input

4 channels 

(isolated between unit and channels)

, HIGH/LOW range switching

Input resistance: 100 kΩ or higher 

(HIGH range)

, 30 kΩ or higher 

(LOW range)

Output (H) detection

170 to 250 V AC, ±DC 70 to 250 V 

(HIGH range)

60 to 150 V AC, ±DC 20 to 150 V 

(LOW range)

Output (L) detection

0 to 30 V AC, ±DC 0 to 43 V 

(HIGH range)

0 to 10 V AC, ±DC 0 to 15 V 

(LOW range)

Response time

Rising edge 1 ms max., falling edge 3 ms max. 

(with HIGH range at 200 V 

DC, LOW range at 100 V DC)

Max. allowable input

250 Vrms 

(HIGH range)

, 150 Vrms 

(LOW range) (the maximum voltage that can 

be applied across input pins without damage)

Cable length and mass:

 Main unit cable 1.5 m (4.92 ft), input section cable 1 m 

(3.28 ft), approx. 320 g (11.3 oz)

Note: The unit-side plug of the 

MR9321-01

 is different from the 

MR9321

.

Cable length and mass:

 70 cm (2.30 ft), Output side: 1.5 m (4.92 ft), 170g (6.0 oz)

DIFFERENTIAL PROBE P9000

(Accuracy guaranteed for 1 year, Post-adjustment accuracy 

guaranteed for 1 year)

Measurement modes

P9000-01: For waveform monitor output, Frequency properties: DC to 100 kHz 

-3 dB

P9000-02: Switches between waveform monitor output/AC effective value output

  Wave mode frequency properties: DC to 100 kHz -3 dB, RMS mode frequency 

properties: 30 Hz to 10 kHz, Response time: Rise 300 ms, Fall 600 ms

Division ratio

Switches between 1000:1, 100:1

DC output accuracy

±0.5 % f.s. 

(f.s. = 1.0 V, division ratio 1000:1), (f.s. = 3.5 V, division ratio 100:1)

Effective value mea-

surement accuracy

±1 % f.s. 

(30 Hz to less than 1 kHz, sine wave)

, ±3 % f.s. 

(1 kHz to 10 kHz, 

sine wave)

Input resistance/capacity

H-L: 10.5 MΩ, 5 pF or less (at 100 kHz)

Maximum input voltage

1000 V AC, DC

Maximum rated volt-

age to ground

1000 V AC, DC (CAT III)

Operating 

temperature range

-40°C to 80°C (-40°F to 176°F)

Power supply

(1) AC adapter Z1008 

(100 to 240 V AC, 50/60 Hz)

, 6 VA 

(including AC 

adapter)

, 0.9 VA 

(main unit only)

(2) USB bus power 

(5 V DC, USB-microB terminal)

, 0.8 VA

(3) External power source 2.7 V to 15 V DC, 1 VA

Accessories

Instruction manual ×1, Alligator clip ×2, Carrying case ×1

WAVE PROCESSOR 9335

Distribution media

One CD-R

Operating 

environment

Computer running under Windows 8/7 (32/64-bit), Vista (32-bit), XP, 2000

Display functions

Waveform display,  X-Y display,  Digital value display,  Cursor function,  

Scroll function,  Maximum number of channels 

(32 channels analog, 32 

channels logic)

,  Gauge display 

(time, voltage axes)

,  Graphical display

File loading

Readable data formats 

(.MEM, .REC, .RMS, .POW)

,  

Maximum loadable file size: Maximum file size that can be saved by a 

given device 

(file size may be limited depending on the computer configuration)

Data conversion

Conversion to

 CSV format,  Tab delimited,  Space delimited,  Data culling 

(simple),  Convert for specified channel,  Batch conversion of multiple 

files

Print functions

Printing image file output 

(expanded META type, “.EMF”)

,  

Supported printer: usable on any printer supported by operating system

Print formatting: 

(1 up, 2-to-16 up, 2-to-16 rows, X-Y 1-to-4 up,  preview,  hard copy)

Other

Parameter calculation,  Search,  Clipboard copy,  Launching of other appli-

cations

Содержание MR8880

Страница 1: ...te settings The re corder can be easily configured to measure voltage drops and power outages Maximum 600V AC DC input no need for a differential probe 4 completely isolated channels let you simultane...

Страница 2: ...22 F Even when operating on battery power the MR8880 can take measurements from 0 C 32 F to 40 C 104 F n Rugged damage resistant design features standard side protectors that guard the instrument s ca...

Страница 3: ...the channel you wish to use Select the power line voltage Select the frequency Select the threshold Select the high speed or real time function The auto range settings can be enabled when using the hi...

Страница 4: ...S s 500 s 8m 20 s 100ms DIV 1 kS s 1 ms 16m 40 s The maximum recording length is fixed regardless of the number of channels in use Measure RMS value fluctuations for a power line over an extended peri...

Страница 5: ...aking measurements in an outdoor setting The dis play is even engineered for easy viewing in the presence of reflections a What if there s no power available in the vehicle being tested A high capacit...

Страница 6: ...annels isolated input and outputs logic has common GND Maximum sampling rate 1Msamples s 1 s cycle all channels simultaneously Memory capacity 14bit 1 M words ch 1 word 2 bytes not expandible External...

Страница 7: ...en and 8 k or lower short Response speed 500ns or lower Max allowable input 0 to 50V DC the maximum voltage that can be applied across input pins without damage LOGIC PROBE MR9321 01 Function Detectio...

Страница 8: ...0 02 Select between WAVE RMS mode For the Memory HiCorder series input up to 1kV AC DC AC ADAPTER Z1008 100 to 240 V AC Voltage to ground is within this product s specifications Separate power source...

Отзывы: