47
Adjust the Zero Position (Zero-Adjust)
2.7 Adjust the Zero Position (Zero-Adjust)
This procedure compensates for module differences and sets the reference potential of the instrument to
0 V.
The compensation procedure is performed for all channels and ranges.
Before Performing Zero-Adjust
•
Before performing this procedure, allow the equipment to warm up for about 30 minutes after the power
is turned on so the internal temperature of the modules can stabilize.
•
Note that zero-adjust cannot be performed during measurement.
•
Key operations are not accepted during zero-adjust.
•
The time required for zero-adjust varies depending on the type and number of modules installed in the
instrument. (The procedure may take several seconds.)
Procedure
Press the
CHAN
key to open the Channel screen, and then select the
[Unit List]
sheet.
1
Move the cursor to
[Zero-Adjust]
.
2
Select
[Exec Zero-Adjust]
.
The zero-adjust procedure is carried out.
Zero-adjustment has no effect on the 8969 and U8969 Strain Unit.
(Perform zero-adjust using auto-balance. (p. 168))
Perform zero-adjust in the following cases.
•
When a module was changed.
•
When power was turned off and on again.
•
When settings were initialized (system reset).
•
When DC/RMS is switched at the 8971 Current Unit and the 8972 DC/RMS Unit
•
When the ambient temperature has changed significantly.
Zero-position drift* may occur.
* Drift:
This refers to spurious output caused by a shift in the operating point of an operational amplifier. Drift can
occur due to changes in temperature and due to component aging over a period of use.
2
Measurement Preparation
Содержание MR8827
Страница 19: ...14 Operation Precautions ...
Страница 81: ...76 Start and Stop Measurement ...
Страница 111: ...106 Manage Files ...
Страница 125: ...120 Miscellaneous Printing Functions ...
Страница 143: ...138 View Block Waveforms ...
Страница 191: ...186 Setting Output Waveform Parameters ...
Страница 291: ...286 FFT Analysis Modes Measurable Ranges With Octave Analysis 1 1 OCT 1 3 OCT ...
Страница 292: ...287 FFT Analysis Modes 1 1 OCT 1 3 OCT 13 FFT Function ...
Страница 293: ...288 FFT Analysis Modes 1 1 OCT 1 3 OCT ...
Страница 295: ...290 FFT Analysis Modes ...
Страница 309: ...304 Editor Command Details ...
Страница 387: ...382 Module Specifications ...
Страница 405: ...400 Dispose of the Instrument Lithium Battery Removal ...
Страница 431: ...A26 FFT Definitions ...
Страница 436: ......