168
Detailed Settings for Modules
8.10.4 Setting Model 8969 and U8969 Strain Unit
The Model 8969 or U8969 Strain Unit can perform auto-balance.
Executing the auto-balance adjusts the reference output level of the transducer to the specified zero
position.
This function can be used for Model 8969 or U8969 Strain Unit only.
The instrument describes Model U8969 as “8969.”
Before Executing Auto-balance
• Turn power on and wait 30 minutes to allow the internal temperature of the module
to stabilize.
• After connecting the strain gauge transducer to the module, execute the auto-
balance without any input including distortion.
• Auto-balance cannot execute during measurement.
• Key operations are not accepted while auto-balance is executing.
To Execute Auto-balance
Refer to “To Open [Each Ch] and Make a Channel Selection” (p. 164).
Setting Item:
[Range]
Auto Bal All Chs
Auto-balance will be executed for all channels where a Model 8969 or
U8969 Strain Unit is installed.
Auto Bal Ch 1
Auto-balance will be executed for the currently selected channel.
Auto-balance can also be executed from the Channel settings window (Analog sheet)
(if the range of a channel with an installed Model 8969 Strain Unit is selected).
To Open the Channel settings window ([Analog] sheet)
Refer to “3.4.2 Choose Analog Channel Settings” (p. 63).
In the following cases, auto-balance should be executed again.
• After changing the vertical axis (strain axis) range
• After a module has been removed or inserted
• After the strain gauge transducer has been replaced
• After power has been turned off and on
• After performing a system reset
•
When ambient temperature has changed significantly (the zero position may drift)
If “Warning: Auto balance failed.” appears
The channel on which auto-balance failed is displayed.
Verify the following, and execute again:
• Is the strain gauge transducer not subject to any load?
(Make sure that it is not subject to vibration, etc.)
• Is the strain gauge transducer connected correctly?
Содержание MR8827
Страница 19: ...14 Operation Precautions ...
Страница 81: ...76 Start and Stop Measurement ...
Страница 111: ...106 Manage Files ...
Страница 125: ...120 Miscellaneous Printing Functions ...
Страница 143: ...138 View Block Waveforms ...
Страница 191: ...186 Setting Output Waveform Parameters ...
Страница 291: ...286 FFT Analysis Modes Measurable Ranges With Octave Analysis 1 1 OCT 1 3 OCT ...
Страница 292: ...287 FFT Analysis Modes 1 1 OCT 1 3 OCT 13 FFT Function ...
Страница 293: ...288 FFT Analysis Modes 1 1 OCT 1 3 OCT ...
Страница 295: ...290 FFT Analysis Modes ...
Страница 309: ...304 Editor Command Details ...
Страница 387: ...382 Module Specifications ...
Страница 405: ...400 Dispose of the Instrument Lithium Battery Removal ...
Страница 431: ...A26 FFT Definitions ...
Страница 436: ......