63
Select Input Channels
3.4.2
Choose Analog Channel Settings
For information about specific settings for Model 8967 Temp Units, 8969 Strain Unit, U8969 Strain Unit,
8970 Freq Unit, 8971 Current Unit, 8972 DC/RMS Unit, MR8990 Digital Voltmeter Unit, and U8974 High
Voltage Unit, refer to “8.10 Detailed Settings for Modules” (p. 164).
Procedure
Press the
DISP
key to open the Waveform screen, and the press the
CH.SET
key to open the channel settings
window (
[Analog]
sheet)
1
Move the cursor to the channel for which
you want to change the settings.
1.
2.
3. 4. 5.
6.
7.
2
Select settings with the
F
keys.
To copy the setting to another channel
Refer to “8.9 Copy Settings to Other Channels (Copy Function)” (p. 163).
1. Waveform
Display Color
Specifies the color in which the waveform of the selected channel is
displayed. You can also select the same color as other channels.
Off
The waveform is not displayed. If the auto-save setting for
[Save
Channel]
is
[Disp Ch]
, data for this channel will not be saved.
Refer to “Select the channels to save.” (p. 84).
On
The waveform is displayed. Select the display color with the
F
keys
[
↑
]
,
[
↓
]
.
All On-Off
Switches the waveform display of all channels to all On or all OFF
2. Vertical Axis
(Voltage Axis)
Range
Sets the vertical axis (voltage axis) range for each channel. The value set
here is the voltage of one increment on the vertical axis.
For information on the full-scale value for various modules, refer to the table
in the section “6. Zero Position” (p. 64).
When the Variable function is
[On]
but variable auto-adjustment is
[Off]
, the
size of the waveform on the screen will not change even if the vertical axis
(voltage) range is changed.
When an over-range condition has occurred
Change the vertical axis (voltage axis) range to a lower sensitivity setting.
3. Coupling
Set the input signal coupling method. Normally, DC coupling should be
selected.
DC (V)
Both DC components and AC components of the input signal will be
measured. (
Default setting)
AC (
V
)
Only AC components of the input signal will be measured. DC
components will be blocked.
GND ( )
Input is shorted to ground. (Allows checking the zero position.)
3
Measurement
Содержание MR8827
Страница 19: ...14 Operation Precautions ...
Страница 81: ...76 Start and Stop Measurement ...
Страница 111: ...106 Manage Files ...
Страница 125: ...120 Miscellaneous Printing Functions ...
Страница 143: ...138 View Block Waveforms ...
Страница 191: ...186 Setting Output Waveform Parameters ...
Страница 291: ...286 FFT Analysis Modes Measurable Ranges With Octave Analysis 1 1 OCT 1 3 OCT ...
Страница 292: ...287 FFT Analysis Modes 1 1 OCT 1 3 OCT 13 FFT Function ...
Страница 293: ...288 FFT Analysis Modes 1 1 OCT 1 3 OCT ...
Страница 295: ...290 FFT Analysis Modes ...
Страница 309: ...304 Editor Command Details ...
Страница 387: ...382 Module Specifications ...
Страница 405: ...400 Dispose of the Instrument Lithium Battery Removal ...
Страница 431: ...A26 FFT Definitions ...
Страница 436: ......