169
Detailed Settings for Modules
8.10.5 Setting Model 8970 Freq Unit
When the standard logic display (LA, LB, LC, LD, LE, LF, LG, and LH) is on, 8970 Freq Units installed
as unit 1, 2, 9, or 10 cannot be used.
Refer to “To Open [Each Ch] and Make a Channel Selection” (p. 164).
Mode
Changes the measurement mode.
Frequency
Measure the frequency of the measurement waveform (Hz hertz)
(Default setting)
RPM
Measure the number of rotations of the measurement target
(r/min: rotations/minutes)
Power
Frequency
Measure the power frequency variation (Hz hertz)
Count
Add up the number of input pulses
Duty
Measure the duty rate of the measurement waveform (% percent)
Pulse width Measure the pulse width (s second)
Pulses with rises during dead time (calculation) (25 kHz or higher) cannot be measured.
Ignored
Calculation (40 μs)
Waveform
loaded
Waveform
loaded
VRange (Input
voltage)
Set the maximum level for the input signal.
±10 V
(Default setting)
, ±20 V, ±50 V, ±100 V, ±200 V, ±400 V
Select the value using the
F
keys.
Threshold value
• When the measurement waveform exceeds the threshold value, the measurement value is
acquired based on the time interval and the number of times the threshold was exceeded.
• The threshold value upper and lower limits and increases and decreases in width
depend on the input voltage ([VRange]) setting.
• While setting the threshold, the voltage level is displayed on the level monitor.
• Select the value using the
F
keys.
To prevent measurement errors due to noise, the threshold has a hysteresis of approximately 3%
versus the input voltage.
(When
[VRange]
is
[ ±10V]
, it is around ±0.3 V.)
Set a threshold in excess of the hysteresis width versus the voltage peak.
8
Utility Functions
Содержание MR8827
Страница 19: ...14 Operation Precautions ...
Страница 81: ...76 Start and Stop Measurement ...
Страница 111: ...106 Manage Files ...
Страница 125: ...120 Miscellaneous Printing Functions ...
Страница 143: ...138 View Block Waveforms ...
Страница 191: ...186 Setting Output Waveform Parameters ...
Страница 291: ...286 FFT Analysis Modes Measurable Ranges With Octave Analysis 1 1 OCT 1 3 OCT ...
Страница 292: ...287 FFT Analysis Modes 1 1 OCT 1 3 OCT 13 FFT Function ...
Страница 293: ...288 FFT Analysis Modes 1 1 OCT 1 3 OCT ...
Страница 295: ...290 FFT Analysis Modes ...
Страница 309: ...304 Editor Command Details ...
Страница 387: ...382 Module Specifications ...
Страница 405: ...400 Dispose of the Instrument Lithium Battery Removal ...
Страница 431: ...A26 FFT Definitions ...
Страница 436: ......