4.1 Setting Criteria to Start and Stop Recording
70
An input signal level “window” within which recording
will (or will not) occur can be defined by upper and
lower threshold levels. You can select whether record-
ing starts or stops when the input signal level enters
(In) or exits (Out) of this window.
Upper and lower thresholds can be verified in the level
monitor or on the Waveform screen. When scaling is
enabled, the scaled threshold values are displayed.
Setting Lower and Upper Trigger Thresholds (Window Trigger)
T
T
T
In
Out
T
Upper/Lower Threshold Ranges and Resolution
Channels
Input
Range
Resolution
Default setting
Upper
Lower
CH1 - CH10
Voltage
100mV
25
V
2 mV
-2 mV
1V
250
V
20 mV
-20 mV
10V
2.5 mV
0.2 V
-0.2 V
20V
5 mV
0.4 V
-0.4 V
100V
25 mV
2 V
-2 V
1-5V
2.5 mV
0.2 V
-0.2 V
Tc
0.5
C
40
C
-40
C
P1 - P4
Count
1 c
80
0
Revolve
1 r/s
80
0
1
Select
[Window]
as the
[Cond]
setting option
for the start or stop trigger.
2
Select inside
[In]
or outside
[Out]
.
Setting options: (
default setting)
3
Enter the upper threshold value for the
[Upper]
setting.
4
Enter the lower threshold value for the
[Lower]
setting.
In
Trigger when the input signal level enters the
window defined by the upper/lower thresholds.
Out
Trigger when the input signal level exits the
window defined by the upper/lower thresholds.
1
2
3
4
When scaling is enabled,
the scaled value is also dis-
played.
Содержание LR8431
Страница 1: ...EN LR8431 20 MEMORY HiLOGGER Instruction Manual Aug 2015 Revised edition 3 LR8431B980 03 15 08H...
Страница 2: ......
Страница 20: ...Operating Precautions 12...
Страница 44: ...2 5 Zero Adjustment 36...
Страница 74: ...3 7 Viewing All Channel Settings in a List 66...
Страница 88: ...4 5 About Recording Operation 80...
Страница 120: ...6 6 Transferring Data to a Computer USB Drive Mode 112...
Страница 126: ...7 2 Numerical Value Calculation Expressions 118...
Страница 136: ...8 3 Making System Settings 128...
Страница 144: ...9 5 Synchronous Measurements with Multiple Instruments 136...
Страница 154: ...Chapter 10 Specifications 146...
Страница 198: ...Appendix 11 Pulse Input Circuit Diagram A40...
Страница 202: ...Index Index 4...
Страница 203: ......
Страница 204: ...EN LR8431 20 MEMORY HiLOGGER Instruction Manual Aug 2015 Revised edition 3 LR8431B980 03 15 08H...