2.5 Zero Adjustment
35
3
2
Chapter 2 M
eas
ureme
n
t Prep
a
rat
ions
Zero adjustment corrects for voltage offset at the input terminals, so that Mem-
ory HiLogger measurements are relative to zero volts.
Perform the zero adjustment procedure when using the instrument in a loca-
tion where the ambient temperature is markedly different.
2.5 Zero Adjustment
Any zero-adjust setting is cleared upon system reset (p. 125).
Turn on the power and wait for 30 minutes to stabilize the internal
temperature of the Memory HiLogger.
With the Waveform or [CH] screen displayed, press and hold these
keys simultaneously for one second.
The Zero Adjust dialog appears.
From the Setting screen
Select
[Zero Adjust]
.
Apply
1
2
Zero adjustment is executed.
(simulta-
neously)
Содержание LR8431
Страница 1: ...EN LR8431 20 MEMORY HiLOGGER Instruction Manual Aug 2015 Revised edition 3 LR8431B980 03 15 08H...
Страница 2: ......
Страница 20: ...Operating Precautions 12...
Страница 44: ...2 5 Zero Adjustment 36...
Страница 74: ...3 7 Viewing All Channel Settings in a List 66...
Страница 88: ...4 5 About Recording Operation 80...
Страница 120: ...6 6 Transferring Data to a Computer USB Drive Mode 112...
Страница 126: ...7 2 Numerical Value Calculation Expressions 118...
Страница 136: ...8 3 Making System Settings 128...
Страница 144: ...9 5 Synchronous Measurements with Multiple Instruments 136...
Страница 154: ...Chapter 10 Specifications 146...
Страница 198: ...Appendix 11 Pulse Input Circuit Diagram A40...
Страница 202: ...Index Index 4...
Страница 203: ......
Страница 204: ...EN LR8431 20 MEMORY HiLOGGER Instruction Manual Aug 2015 Revised edition 3 LR8431B980 03 15 08H...