Procedure

1

Make sure that power of the instrument to connect 

the L2020 to is off.

2

Connect the L2020 to 

the input terminal of the 

instrument.

Red

Black

Align the similarly 

colored 

 marks.

3

Perform zero adjustment.

Be sure to use a zero 

adjustment board which is 

the accessory or option of 

connectable instrument.

(See the instruction manual for 

details of connectable devices.)

SOURCE

SENSE

Zero adjustment board

4

Connect the L2020 to a sample.

Part Names

(1)

(1)

(3)

(5)

(4)

(2)

(1)

Protective cap

Remove this cap before use.

Reattach the cap when not using the lead.

(2)

Conductive contact pin

(3)

Grip

(4)

SOURCE connector

(5)

SENSE connector

Enlarged view of pin

13.5

φ

1.27

φ

2.9

φ

2.7

Unit: mm

1

Specification

General Specification

Operating 

environment 

Indoors, Pollution Degree 2, altitude up 

to 2000 m (6562 ft.)

Operating temperature 

and humidity

0°C to 40°C (32°F to 104°F),  

80% RH or less (no condensation)

Storage temperature 

and humidity

-10°C to 50°C (14°F to 122°F), 

80% RH or less (no condensation)

Dimensions

Approx. 1950 mm  (76.77″)

Mass

Approx. 210 g  (7.4 oz.)

Accessory

Instruction manual

Option

Model 9465-90 Tip Pin

Basic Specification

Rated current

2 A AC/DC continuous

Maximum rated 

voltage to earth

Effective value 30 V, Peak value 42.4 V, 

Direct current 60 V

Measurement 

connector

SOURCE Hi, SOURCE Lo, SENSE Hi, 

SENSE Lo

No GUARD connector

Replacing the Pin Tip (Option)

The conductive contact pin is replaceable. Replace the pin 

with a new one if it is broken or worn. Purchase the 9465-90 

Pin Tip, which includes the conductor contact pin and the pin 

base (plastic portion).

1

Turn off the power of the instrument and remove 

the L2020.

2

Rotate the grip to loosen 

it.

Beware of injury.

The tip has a sharp point.

3

Remove the pin tip.

9465-90 Pin Tip

4

Replace the pin tip to a 

new one.

9465-90 Pin Tip

5

Rotate the grip to fasten 

it.

Be sure to fasten it tightly.

6

To avoid broken wires and contact failures, check 

that the cable is firmly held

7

Check the performance.

Measure an object with a known resistance. Make

 

sure that 

the measured resistance is correct before using the pin type 

lead.

IMPORTANT

Measurement values when using 4-terminal measurement 

(Differences in measurement values due to pin type leads 

used)

Depending on the subject of measurement, such as a lead-acid 

battery, measurement values may vary due to the pin type lead 

used. Since these differences in measurement values are due to 

the shapes and dimensions of the pin type leads used in 4-terminal 

measurement, measurement values taken using any pin type lead 

represent the true values for that pin type lead only.

When judging battery wear using changes in resistance values with 

time, be sure to use pin type leads having the same dimensions.

Reference example: measurement of an MSE-200 valve-regulated 

stationary lead-acid battery

(Resistance values vary according to the materials and structure of 

the terminals of the subject of measurement.)

Pin type lead (Distance between 

the current-impression pin and the 

voltage-measurement pin)

Measurement values

using the BT3554 

Battery Tester

L2020 Pin Type Lead (0.65 mm)

0.538 m

Ω

9772 Pin Type Lead (2.5 mm)

0.490 m

Ω

See the BT3554 Battery Tester Instruction manual for detailed 

technical descriptions.

Explanation

Differences in measurement values are physical phenomena 

resulting from differences in the distances (dimensions) between 

current-impression pins and voltage-measurement pins. The 

greater the battery terminal resistance in comparison to the 

battery's internal resistance, the more marked these differences 

become. The following diagram shows how differences in voltage 

detected result from differences in distance when measuring a 

lead-acid battery.

Model L2020
(Pin distance: 0.65 mm)

Model 9772
(Pin distance: 2.5 mm)

+

+

Lead-acid battery

Lead-acid battery

Equipotential line

Equipotential line

Potential gradient

Potential gradient

V (voltage detected): A>B

A

B

Отзывы: