55
General Specifications
Degree of protection
IP40
Maximum rated
voltage to terminal
600 V AC/DC (Voltage measurement)
600 V AC/1000 V DC (Voltage measurement,
only for IR4055)
Maximum rated
voltage to earth
600 V AC/DC, Measurement Category III,
Anticipated Transient Overvoltage: 6000 V
Dielectric strength
7060 V AC, 50 Hz/60 Hz, between measurement
terminals and chassis, 1 minute, current
sensitivity 1 mA
Power supply
Rated supply voltage: 1.5 V DC × 4
LR6 alkaline battery × 4
Maximum rated power
3 VA
Continuous operating
time (at 23°C, as a
referential)
Bluetooth
®
OFF: Approx. 20 hours
(Comparator off, backlight off, measured with
measurement terminal open at 500 V range)
Bluetooth
®
ON: Approx. 10 hours
(Comparator off, backlight off, measured with
measurement terminal open at 500 V range)
Drop proof
On concrete: 1 m
Fuse (Replacements)
FF0.5 AH/1000 V (70 172 40.0.500: SIBA GmbH)
(Very fast-blow, arc-extinguishing material
included, high breaking capacity)
Dimensions (excluding
protrusions)
Approx. 159W × 177H × 53D mm
(6.26″W × 6.97″D × 2.09″D)
Mass (including
battery, excluding test
lead)
• IR4055, IR4056: Approx. 600 g (21.2 oz.)
• IR4057, IR4058: Approx. 640 g (22.6 oz.)
Accessories
Refer to “Verifying Package Contents” (p. 1)
Options
4
1
2
3
5
6
7
付録