Making Measurements
Making Measurements
Measurement Procedure
You will need (besides the device and accessories):
• IM7580 series Impedance Analyzer ×1
• IM9200 Test Fixture Stand ×1
• IM9905 Calibration Kit ×1
• IM9906 Adapter ×1
• 3.5 mm connector torque wrench ×1 (not included with the device)
• Tweezers ×1 (This is not included with the device. Tweezers which are thinner than the DUT size,
clean, and not magnetized are recommended.)
1
Connect the test head to the
Connect the test head to the
impedance analyzer.
impedance analyzer.
Connect the cable of the test head to the impedance
analyzer.
See the instruction manual
of the impedance analyzer.
Test head
2
Attach the test head and adapter
Attach the test head and adapter
to the stand.
to the stand.
Attach the test head and the IM9906 Adapter to the
IM9200 Test Fixture Stand.
IM9200
IM9906
3
Execute the calibration.
Execute the calibration.
Execute the open calibration, short calibration, and load
calibration.
4
Attach the device.
Attach the device.
Attach the device to the IM9200 Test Fixture Stand.
Device (IM9202)
5
Execute the correction.
Execute the correction.
Execute the electrical length correction, short correction,
and open correction.
(From p. 15)
6
Start the measurement.
Start the measurement.
Connect the DUT to the measurement location and start
the measurement by using the impedance analyzer.
HIOKI IM9202A961-00