5.10 Equivalent Circuit Analysis Function
262
When the comparator is on and a judgment area has been set, the estimated values and judgment results will
be displayed after equivalent circuit estimation. Judgment results can also be acquired using communications
commands or external output (EXT I/O).
2
Making judgments using analysis results
8
Press
to close the setting screen.
7
Setting the equivalent circuit
You can check the set upper and lower limit values
on the Equivalent Circuit Settings screen.
Top row
: Upper limit values
Bottom row : Lower limit values
Display of judgment results
Estimated value is above
upper limit
Pass (meets criteria)
Estimated value is below
lower limit
Changing estimated values manually causes judgment results to be cleared.
The overall judgment result is output with the front panel's LED as well as from EXT I/O pin 14.
See
"Chapter 11 External Control" (p. 363)
However, the judgment content differs depending on whether you are performing analysis manually
or automatically.
For more information, see the following table:
Method of
analysis
Judgment timing
Overall judgment result
MANUAL
When measurement completes
Outputs the area comparator or peak comparator judgment result.
Outputs nothing when the area comparator or peak comparator have
not been configured.
When equivalent circuit analysis is
performed manually
Clears the area comparator or peak comparator judgment results
and outputs the overall judgment result for the equivalent circuit
analysis results.
AUTO
When equivalent circuit analysis is
performed after measurement
completes
Outputs the area comparator or peak comparator judgment results
as well as the overall judgment result for the equivalent circuit
analysis results.
Содержание IM3570
Страница 1: ...IM3570 Instruction Manual IMPEDANCE ANALYZER EN Dec 2018 Revised edition 8 IM3570A981 08 18 12H ...
Страница 2: ......
Страница 8: ...Contents vi ...
Страница 16: ...Operating Precautions 8 ...
Страница 34: ...1 3 Screen Configuration and Operation 26 ...
Страница 138: ...4 5 Setting Application Settings 130 ...
Страница 156: ...5 3 Normal Sweep 148 When the setting is or Procedure Analyzer Initial Screen Press 2 1 Example When ...
Страница 160: ...5 3 Normal Sweep 152 When the setting is Procedure Analyzer Initial Screen Press 2 1 ...
Страница 272: ...5 10 Equivalent Circuit Analysis Function 264 ...
Страница 336: ...9 4 Deleting a Panel 328 ...
Страница 410: ...13 3 About Measurement Times and Measurement Speed 402 ...
Страница 418: ...14 4 Discarding the Instrument 410 ...
Страница 444: ...Appendix 13 Device Compliance Statement A26 ...
Страница 447: ...Index3 Index Vertical Axis Scale 183 Voltage limit 60 78 ...
Страница 448: ...Index4 ...
Страница 455: ......
Страница 456: ......
Страница 457: ......