87
Setting Application Settings
Memory function (saving measurement results)
You can save the measurement results inside the instrument (Up to 32,000 items).This function al
-
lows you to save previously saved measurement results to the USB flash drive and to acquire them
from a computer using the
:MEMory?
communications command.
When using communications commands, the information saved to memory reflects the
:MEASure:VALid
.
For more information about acquiring measurement results stored in memory and how to configure
the
:MEASure:VALid
setting, see the LCR application disc (communications commands).
IMPORTANT
Changing the memory function setting will cause data stored in the instrument’s memory to be
deleted.
Screen display method (For more information, see p. 24):
(Measurement screen)
SET
key>(
SET
screen)
ADVANCED
tab>
MEMORY
key
1
After touching the
OFF
key to disable the
memory function, use the
keys to
set the number of measurement results.
1
2
Settable range: 1 to 32000
The number of measurement results can
only be set when the memory function is
set to
OFF
.
2
Touch the
IN
key or
ON
key.
Measured values in the
instrument’s memory are
deleted once they have
been saved to the USB
flash drive. The measure
-
ment values are saved
to the
MEMORY
folder in
the USB flash drive. The
file name is automatically
assigned from the date
Clears all of the
saved measure-
ment values
from the instru
-
ment memory.
IN
Measured values are stored in memo-
ry only when all parameters judged by
the comparator and BIN measurement
yield a pass result. (If even one HI or
LO comparator result is received or if
the BIN result is OUT OF BINS, the
value will not be saved.)
ON
Saves all measurement values to
memory.
If the comparator and BIN functions have not
been configured, the
IN
and
ON
keys result
in the same operation.
3
Touch the
EXIT
key twice.
Displays the measurement screen.
3
Performing Measurements in LCR Mode
Содержание IM3536
Страница 20: ...16 Operating Precautions ...
Страница 34: ...30 Screen Layout and Operation ...
Страница 140: ...136 Testing the System Self diagnosis ...
Страница 224: ...220 About Measurement Times and Measurement Speed ...
Страница 240: ...236 Discarding the Instrument ...
Страница 254: ...Appx 14 Dimensional Diagram Appx 10 Dimensional Diagram 230 114 15 48 33 33 25 35 119 1 330 1 Unit mm ...
Страница 264: ...Appx 24 Device Compliance Statement ...
Страница 270: ...Test Equipment Depot 800 517 8431 99 Washington Street Melrose MA 02176 TestEquipmentDepot com ...