3.4  RS-232C Connection and Settings (IM3523 only, when connected to the Z3001)

27

Connect the RS-232C cable to the RS-232C connector.
(Recommended cable: 9637 RS-232C cable)

Example: Connecting to a DOS/V PC

Specification: D-sub 9-pin female and female connector, reverse connection

3.4 RS-232C Connection and Settings

(IM3523 only, when connected to the 

Z3001)

Connecting the RS-232C Cable

Male 9-pin D-sub
#4-40 attaching screws

To connect the instrument to a controller (DTE), use a 

crossover cable

 

compatible with the connectors on both the instrument and the controller.
The I/O connector is a DTE (Data Terminal Equipment) configuration.

6  7  8  9

1  2  3  4  5

Connector (D-sub)  

Pin No.

Interchange Circuit 

Name

CCITT

Circuit No.

EIA

Abbreviation

JIS

Abbreviation

Common 

Abbreviation

1

Unused

2

Received Data

104

BB

RD

RxD

3

Transmitted Data

103

BA

SD

TxD

4

Data Terminal Ready

108/2

CD

ER

DTR

5

Signal Ground

102

AB

SG

GND

6

Unused

7

Request to Send

105

CA

RS

RTS

8

Clear to Send

106

CB

CS

CTS

9

Unused

Hardware control will not work properly if you use a cable that has CA(RTS) and CB(CTS)
short-circuited.  

HIOKI IM3570A983-08

Содержание IM3523

Страница 1: ...nstruction Manual EN Dec 2022 Revised edition 8 IM3570A983 08 22 12H IM3523 IM3533 LCR METER IM3533 01 IM3536 IMPEDANCE ANALYZER IM3570 IM7580 CHEMICAL IMPEDANCE ANALYZER IM3590 IM3523A 600329708 HIOK...

Страница 2: ...HIOKI IM3570A983 08...

Страница 3: ...g the USB Cable 24 3 3 GP IB Connection and Settings IM3523 only when connected to the Z3000 25 Connecting the GP IB Cable 25 Setting GP IB 25 3 4 RS 232C Connection and Settings IM3523 only when conn...

Страница 4: ...and Connection 55 LAN Settings 55 Connecting a LAN Cable 60 5 4 GP IB Connection and Settings when connected to the Z3000 61 Connecting the GP IB Cable 61 Setting GP IB 61 5 5 RS 232C Connection and S...

Страница 5: ...site https www hioki com global support download Product registration Register your product in order to receive important product information https www hioki com global support myhioki registration Sa...

Страница 6: ...Press and quickly release the right button of the mouse Double click Quickly click the left button of the mouse twice Drag While holding down the left button of the mouse move the mouse and then rele...

Страница 7: ...When using hardware RTS CTS control When CB CTS is confirmed to be OFF the sending of data is halted When it is confirmed to be ON the sending of data is resumed When using software XON XOFF control...

Страница 8: ...ons RL1 Supports all remote local functions PP0 Parallel poll functions are not supported DC1 Supports all device clear functions DT1 Supports all device trigger functions C0 Controller functions are...

Страница 9: ...isconnecting an interface connector Otherwise an electric shock accident may occur To avoid damage to the instrument do not short circuit the terminal and do not input voltage to the terminal Failure...

Страница 10: ...nterface the icon for the set interface is displayed on the right side of the screen Screen Displayed while Setting Interfaces When RS 232C is set When GP IB is set When USB is set When LAN is set Whe...

Страница 11: ...compatible with the connectors on both the instrument and the controller The I O connector is a DTE Data Terminal Equipment configuration 6 7 8 9 1 2 3 4 5 Connector D sub Pin No Interchange Circuit N...

Страница 12: ...the setting from any of mode and mode only IM3570 Press 2 RS 232C Settings 3 Press to confirm the setting 4 RS 232C Settings Select the terminator setting CR LF CR Select the baud rate setting Select...

Страница 13: ...nector 2 3 GP IB Connection and Settings Connecting the GP IB Cable Setting GP IB Recommended cable 9151 02 GP IB connection cable 2 m Procedure LCR Measurement Screen 1 Interface Settings You can con...

Страница 14: ...IB Connection and Settings 10 Press 2 GPIB Setting 3 GPIB Setting Use or to set the GP IB address Select the terminator setting LF with EOI LF with CR EOI Press to confirm the setting 4 HIOKI IM3570A...

Страница 15: ...ed CD or our web site http www hioki com The USB driver is compatible with the Windows 7 32 bit 64 bit version Windows 8 32 bit 64 bit version Windows 10 32 bit 64 bit version and Windows 11 64 bit ve...

Страница 16: ...ly available USB cable to the USB port of the instrument Connecting the USB Cable 3 USB Setting Select the terminator setting CR LF CR Press to confirm the setting 4 USB cable commercially available p...

Страница 17: ...erent IP address than the computer Subnet Mask 255 255 255 0 Default Gateway OFF 0 0 0 0 2 5 LAN Settings and Connection LAN Settings Make these settings before connecting to a network Changing settin...

Страница 18: ...and Connection 14 Procedure LCR Measurement Screen 1 Interface Settings You can configure the setting from any of mode and mode only IM3570 Press 2 LAN Settings LAN Settings 3 Select the IP address HI...

Страница 19: ...0 255 255 224 000 255 255 255 224 240 000 000 000 255 240 000 000 255 255 240 000 255 255 255 240 248 000 000 000 255 248 000 000 255 255 248 000 255 255 255 248 252 000 000 000 255 252 000 000 255 25...

Страница 20: ...ay does not need to be set for example when connecting the instrument and computer on a one to one basis using a cross ca ble leave this set to OFF Default gateway Settings 8 Use or to set the default...

Страница 21: ...er Settings 10 Use or to set the port number to use for communication commands Press to confirm the setting Settable range 1024 to 65535 LAN Settings 11 Select the terminator setting CR LF CR Press to...

Страница 22: ...ecting a LAN Cable When connecting the instrument to your LAN using a LAN cable of more than 30 m or with the cable laid outdoors take appropriate countermeasures that include install ing a surge prot...

Страница 23: ...ode remote operation state and the keys on the LCD are disabled 2 6 Remote Mode Canceling Remote Mode All of the keys except are disabled Remote Mode State Procedure 2 Press to return to the normal st...

Страница 24: ...2 6 Remote Mode 20 HIOKI IM3570A983 08...

Страница 25: ...ircuit the terminal and do not input voltage to the terminal Failure to fasten the connectors properly may result is sub specification perfor mance or damage to the equipment To avoid damage do not di...

Страница 26: ...USB and LAN inter faces There are the following two communication methods To enable communication the communication condi tions need to be set on the instrument The instrument is communication class...

Страница 27: ...m IM3523A When the instrument is connected to a computer the USB driver is automatically installed Since the OS standard driver is installed it is not necessary to install another driver The USB drive...

Страница 28: ...o the USB port of the instrument 3 Select the terminator setting 4 Connecting the USB Cable Select 1 Select the terminator setting 2 You will return to the measure ment screen USB cable commercially a...

Страница 29: ...GP IB connector 3 3 GP IB Connection and Settings IM3523 only when connected to the Z3000 Connecting the GP IB Cable Setting GP IB 1 Open the SYSTEM screen 2 Select GP IB as the interface Recommended...

Страница 30: ...nected to the Z3000 26 3 Select the terminator setting 4 Set the GP IB address Valid setting range 0 to 30 5 Select 1 LF with CR EOI 2 LF with EOI Select 1 Select the GP IB address setting 2 You will...

Страница 31: ...troller DTE use a crossover cable compatible with the connectors on both the instrument and the controller The I O connector is a DTE Data Terminal Equipment configuration 6 7 8 9 1 2 3 4 5 Connector...

Страница 32: ...the SYSTEM screen 2 Select RS 232C as the interface 3 Select from the following baud rate setting 9600 19200 38400 57600 4 Select the terminator setting Open the SYSTEM screen Select 1 Select RS 232C...

Страница 33: ...d to the Z3001 29 5 Select the handshake setting 6 Select 1 Select the handshake setting 2 OFF No flow control HARD Hardware RTS CTS control XON OFF Software XON XOFF control BOTH Hardware software Yo...

Страница 34: ...erent IP address than the computer Subnet Mask 255 255 255 0 Default Gateway OFF 0 0 0 0 3 5 LAN Settings and Connection IM3523 needs to be connected to the Z3002 LAN Settings Make these settings befo...

Страница 35: ...255 192 000 000 255 255 192 000 255 255 255 192 224 000 000 000 255 224 000 000 255 255 224 000 255 255 255 224 240 000 000 000 255 240 000 000 255 255 240 000 255 255 255 240 248 000 000 000 255 248...

Страница 36: ...he default gateway 2 Set the default gateway to OFF 000 000 000 000 If the default gateway does not need to be set for example when connecting the instrument and computer on a one to one basis using a...

Страница 37: ...ument to a single computer connect the instrument to the computer Connecting a LAN Cable On Performing communication with 100BASE Off Performing communication with 10BASE Flashes when data is being ex...

Страница 38: ...n the mode becomes remote mode remote operation state and the keys on the LCD are disabled 3 6 Remote Mode Remote status Canceling Remote Mode 1 2 Keys other than F1 are disabled You will return to th...

Страница 39: ...ument do not short circuit the terminal and do not input voltage to the terminal Failure to fasten the connectors properly may result is sub specification perfor mance or damage to the equipment To av...

Страница 40: ...et an interface the icon for the set interface is displayed on the right side of the screen Screen Displayed while Setting Interfaces When USB is set When GP IB is set When RS 232C is set When LAN is...

Страница 41: ...ur web site http www hioki com The USB driver is compatible with the Windows 7 32 bit 64 bit version Windows 8 32 bit 64 bit version Windows 10 32 bit 64 bit version and Windows 11 64 bit version oper...

Страница 42: ...ly available USB cable to the USB port of the instrument Connecting the USB Cable 3 USB Setting Select the terminator setting CR LF CR Press to confirm the setting 4 USB cable commercially available p...

Страница 43: ...connector 4 3 GP IB Connection and Settings when connected to the Z3000 Connecting the GP IB Cable Setting GP IB Recommended cable 9151 02 GP IB connection cable 2 m You can configure the setting fro...

Страница 44: ...and Settings when connected to the Z3000 40 Press 2 GPIB Setting 3 GPIB Setting Use or to set the GP IB address Select the terminator setting LF with EOI LF with CR EOI Press to confirm the setting 4...

Страница 45: ...use a crossover cable compatible with the connectors on both the instrument and the controller The I O connector is a DTE Data Terminal Equipment configuration 6 7 8 9 1 2 3 4 5 Connector D sub Pin No...

Страница 46: ...You can configure the setting from any of mode mode and mode Press 2 RS 232C Settings 3 Press to confirm the setting 4 RS 232C Settings Select the handshake setting No flow control Hardware RTS CTS co...

Страница 47: ...erent IP address than the computer Subnet Mask 255 255 255 0 Default Gateway OFF 0 0 0 0 4 5 LAN Settings and Connection when connected to the Z3002 LAN Settings Make these settings before connecting...

Страница 48: ...ion when connected to the Z3002 44 You can configure the setting from any of mode mode and mode Procedure LCR Measurement Screen 1 Interface Settings Press 2 LAN Settings LAN Settings 3 Select the IP...

Страница 49: ...255 224 000 000 255 255 224 000 255 255 255 224 240 000 000 000 255 240 000 000 255 255 240 000 255 255 255 240 248 000 000 000 255 248 000 000 255 255 248 000 255 255 255 248 252 000 000 000 255 252...

Страница 50: ...default gateway does not need to be set for example when connecting the instrument and computer on a one to one basis using a cross ca ble leave this set to OFF Default gateway Settings 8 Use or to s...

Страница 51: ...2 47 Port number Settings 10 Use or to set the port number to use for communication commands Press to confirm the setting Settable range 1024 to 65535 LAN Settings 11 Select the terminator setting CR...

Страница 52: ...with cross over adapter up to 100 m Hioki 9642 LAN Cable option When connecting the instrument to a single computer connect the instrument to the computer Connecting a LAN Cable On Performing communi...

Страница 53: ...ode remote operation state and the keys on the LCD are disabled 4 6 Remote Mode Canceling Remote Mode All of the keys except are disabled Remote Mode State Procedure 2 Press to return to the normal st...

Страница 54: ...4 6 Remote Mode 50 HIOKI IM3570A983 08...

Страница 55: ...erminal and do not input voltage to the terminal Failure to fasten the connectors properly may result is sub specification perfor mance or damage to the equipment To avoid damage do not disconnect the...

Страница 56: ...When you set an interface the icon for the set interface is displayed on the right side of the screen Screen Displayed while Setting Interfaces When USB is set When GP IB is set When RS 232C is set Wh...

Страница 57: ...an be downloaded from the bundled CD or our web site http www hioki com The USB driver is compatible with the Windows 7 32 bit 64 bit version Windows 8 32 bit 64 bit version Windows 10 32 bit 64 bit v...

Страница 58: ...ly available USB cable to the USB port of the instrument Connecting the USB Cable 3 USB Setting Select the terminator setting CR LF CR Press to confirm the setting 4 USB cable commercially available p...

Страница 59: ...erent IP address than the computer Subnet Mask 255 255 255 0 Default Gateway OFF 0 0 0 0 5 3 LAN Settings and Connection LAN Settings Make these settings before connecting to a network Changing settin...

Страница 60: ...tings and Connection 56 You can configure the setting from any of mode and mode Procedure LCR Measurement Screen 1 Interface Settings Press 2 LAN Settings LAN Settings 3 Select the IP address HIOKI IM...

Страница 61: ...0 255 255 224 000 255 255 255 224 240 000 000 000 255 240 000 000 255 255 240 000 255 255 255 240 248 000 000 000 255 248 000 000 255 255 248 000 255 255 255 248 252 000 000 000 255 252 000 000 255 25...

Страница 62: ...way does not need to be set for example when connecting the instrument and computer on a one to one basis using a cross ca ble leave this set to OFF Default gateway Settings 8 Use or to set the defaul...

Страница 63: ...er Settings 10 Use or to set the port number to use for communication commands Press to confirm the setting Settable range 1024 to 65535 LAN Settings 11 Select the terminator setting CR LF CR Press to...

Страница 64: ...e with cross over adapter up to 100 m Hioki 9642 LAN Cable option When connecting the instrument to a single computer connect the instrument to the computer Connecting a LAN Cable On Link up Off Link...

Страница 65: ...IB connector 5 4 GP IB Connection and Settings when connected to the Z3000 Connecting the GP IB Cable Setting GP IB Recommended cable 9151 02 GP IB connection cable 2 m You can configure the setting f...

Страница 66: ...and Settings when connected to the Z3000 62 Press 2 GPIB Setting 3 GPIB Setting Use or to set the GP IB address Select the terminator setting LF with EOI LF with CR EOI Press to confirm the setting 4...

Страница 67: ...sub 4 40 attaching screws To connect the instrument to a controller DTE use a crossover cable compatible with the connectors on both the instrument and the controller The I O connector is a DTE Data...

Страница 68: ...en 1 Interface Settings You can configure the setting from any of mode and mode Press 2 RS 232C Settings 3 Press to confirm the setting 4 RS 232C Settings Select the handshake setting No flow control...

Страница 69: ...ode remote operation state and the keys on the LCD are disabled 5 6 Remote Mode Canceling Remote Mode All of the keys except are disabled Remote Mode State Procedure 2 Press to return to the normal st...

Страница 70: ...5 6 Remote Mode 66 HIOKI IM3570A983 08...

Страница 71: ...irtual COM Port Device Manager starts The procedure to start Device Manager differs depending on the version of the Windows operat ing system For details refer to Help of the operating system Check Ch...

Страница 72: ...Appendix 1 Checking the USB Virtual COM Port A2 HIOKI IM3570A983 08...

Страница 73: ...HIOKI IM3570A983 08...

Страница 74: ...HIOKI IM3570A983 08...

Страница 75: ...HIOKI IM3570A983 08...

Страница 76: ...www hioki com HIOKI IM3570A983 08...

Отзывы: