3.2 Input, Output, and Measurement Specifications
59
AC current measurement
Accuracy
guarantee
conditions
Sine wave input
Measurement
method
Digital sampling method (true RMS measurement)
Crest factor
5.0 or less (for the 60 A range, 1.7 or less)
Conductor
position effects
±0.5% rdg or less (using the center of the sensor as the reference,
in all positions)
Magnetic field
interference
10 mA or less in an external magnetic field of 400 A/m at 50 Hz /60 Hz
AC
Maximum
allowable current
100 A AC continuous, 200 A AC for 2 minutes (50 Hz/60 Hz)
For frequency derating characteristics during continuous input, see
the following diagram:
Effective
measuring range
0.05 mA to 60.0 A
Zero display
range
Less than 0.05 mA
Overrange
Greater than 60.0 A
Measurement
response time
Filter off/ Filter on: 1 s ±0.5
s or less
Input-current frequency [Hz]
M
a
xi
m
u
m
al
lo
wabl
e
c
u
rrent
[A
]
HIOKI FT6380B981-01
www.calcert.com
1.800.544.2843
0
5
10
15
20
25
30