DT4281,DT4282 DIGITAL MULTIMETER Remote Operation Manual
P6/14
5. Detail of Commands
5.1. Description of Commands
The DT4281,DT4282 only accepts the UPPER CASE command. (Except Unit etc.)
Table.4
Command
Explanation
QPID
Query the Meter model.
Syntax QPID
Response “DT4281” or “DT4282”
*IDN
Query the Meter identification.
Syntax *IDN?
Response <data×3>
<data> maker name, model number, serial, version
Example; “HIOKI,DT4281,121107517,Ver 1.00”
:SYST:RST
Put the meter to reset.
Syntax :SYST:RST
Response “OK”
:SYST:LLO
Put the meter into the local lockout state.
Syntax :SYST:LLO
Response “OK”
:SYST:GTL
Put the meter into the local state.
Syntax :SYST:GTL
Response “OK”
:CONF?
:CONF2?
Query the function of the display.
Syntax :CONF? [:CONF2?]
Response <data×2>
<data> function,range
Example; “ACV, 600m”
5.2 Combination response string function and range
:CONF
Configure the range.
Syntax :CONF <data×2>
<data> function,range
Example; “:CONF RES, 60k”
5.2 Combination response string function and range
Response “OK” or “CMD ERR” or “EXE ERR”