20
Wireless Communications Function
Z3210-to-Excel
®
direct data entry function (Excel
®
direct input
function, HID function)
Concurrent use of GENNECT Cross (p. 17) and the HID function is not available.
The human interface device (HID) profile, with which the Z3210 Wireless Adapter is equipped, is a
profile same as that wireless keyboards use.
HID ON
Preparatory to data entry, open an Excel
®
file on your mobile device or computer
and choose a cell. When the instrument’s display freezes, the measured values
are entered on the cells. The use of this function with the auto hold function
enabled comes in handy. (p. 9)
HID OFF
When you wish to use GENNECT Cross, disable the HID function.
The setting whether the HID function has been enabled or disabled is not saved in the instrument
but in the Z3210.
Confirming and toggling the HID setting
1
Remove the instrument from the object under measurement and turn off the instrument.
2
Connect the Z3210 Wireless Adapter (option) to the instrument.
3
Turn the power on while holding down the
and RANGE keys simultaneously.
The HID setting saved in the Z3210 is displayed.
When the HID function is enabled
When the HID function is disabled
If you do not wish to change the HID setting, press the power key to turn off the
instrument.
4
Press the MAX/MIN or RANGE key to enable or disable the HID function.
The
HOLD
key blinks. Every time you press the key, the HID function switches on and off.
HIOKI CM4001A963-00