Zero adjustment has failed
The buzzer sounds and
[FAIL]
appears. Then, the measurement screen
appears.
Zero adjustment has failed
When zero adjustment cannot be performed, the measured value before
zero adjustment already exceeds ±3% of the full scale of each range or the
instrument is in a measurement error state. Perform zero adjustment with
the correct wire connection again. If the resistance is too high (e.g., due to a
self-made cable), zero adjustment cannot be performed. In such a case, try
to minimize the wiring resistance.
Clearing zero adjustment
When the power is off, while holding the
[0 ADJ]
key, press the
[POWER]
key to clear the zero adjustment for all ranges.
IMPORTANT
If zero adjustment fails, the zero adjustment is cleared for the current range.
Содержание CKT3548
Страница 1: ...CKT3548 User s Manual RESISTANCE METER Changzhou Chuangkai Electronic Co Ltd Edition V1 0...
Страница 19: ...Changzhou Chuangkai Electronic Co Ltd Edition V1 0 19 1 5 Dimension...
Страница 22: ...Changzhou Chuangkai Electronic Co Ltd Edition V1 0 22 Lithium battery replacement...
Страница 40: ...Changzhou Chuangkai Electronic Co Ltd Edition V1 0 40...
Страница 51: ...Changzhou Chuangkai Electronic Co Ltd Edition V1 0 51 Judging based on upper and lower limit values ABS mode...
Страница 56: ...Changzhou Chuangkai Electronic Co Ltd Edition V1 0 56 Measurement screen...
Страница 57: ...Changzhou Chuangkai Electronic Co Ltd Edition V1 0 57...
Страница 71: ...Changzhou Chuangkai Electronic Co Ltd Edition V1 0 71 Clearing an entire block of saved data Clearing all...
Страница 81: ...Changzhou Chuangkai Electronic Co Ltd Edition V1 0 81...
Страница 84: ...Changzhou Chuangkai Electronic Co Ltd Edition V1 0 84...