Hope Electronics Technology
Manual
Edition
V1.0
49
oF
Hi
-oF
Lo
--------
No judgment
If the power is turned off during a setting process, any setting changes are
lost and the previous values remain valid. To apply the changes, press the
[ENTER]
key
ABS (absolute value judgment) mode
Set the upper and lower limit values for judgment, as absolute values.
Example:
Upper limit value……………..100.00 mΩ
Lower limit value……………..80.00 mΩ
Upper limit value …………….100.00 mΩ
REF% (relative value judgment) mode
Set the allowable % of a reference value to determine the upper and lower
limit
values for judgment. In REF% mode, the upper and lower limit values
cannot be
set separately.
Example:
Reference value………………………….12.000kΩ
Upper and lower limit values…………..±1.00%
Turning the comparator function ON/OFF