The ERR output signal indicates the occurrence of measurement fault conditions
(such as open test leads, or a bad contact).
There are two ERR output methods.
Synchronized
with EOM Out
-
put (SYNC)
Measurement faults detected while measuring (not while awaiting trigger or
during delay or calculation intervals), are indicated by ERR output synchronous
with EOM output (the end-of-measurement signal).
ERR Output Low (On): A measurement fault has prevented correct measurement
ERR Output High (Off): Correct measurement obtained (OF or -OF: Out-of-
range cases are included)
Asynchronous
with EOM Out
-
put (ASYNC)
Measurement faults (test lead connection conditions) are output in real time. The
output is asynchronous with the TRIG signal and EOM output.
ERR Output Low (On): Measurement fault condition (open test leads, or a bad
contact)
ERR Output High (Off): Test lead connections are normal
ERR Output
HIOKI BT3562A981-09
Содержание BT3562
Страница 2: ...HIOKI BT3562A981 09...
Страница 24: ...1 5 Measurement Flowchart 18 HIOKI BT3562A981 09...
Страница 84: ...4 13 Reset Function 78 HIOKI BT3562A981 09...
Страница 93: ...5 4 Internal Circuitry 87 5 Chapter 5 External Control EXT I O 5 4 Internal Circuitry HIOKI BT3562A981 09...
Страница 102: ...6 3 Printing 96 HIOKI BT3562A981 09...
Страница 186: ...8 8 Sample Programs 180 HIOKI BT3562A981 09...
Страница 198: ...10 3 Error Display 192 HIOKI BT3562A981 09...
Страница 200: ...194 HIOKI BT3562A981 09...
Страница 220: ...Appendix 11 Dimensional Diagram A20 HIOKI BT3562A981 09...
Страница 224: ...Index HIOKI BT3562A981 09...
Страница 225: ...HIOKI BT3562A981 09...
Страница 226: ...HIOKI BT3562A981 09...
Страница 227: ...HIOKI BT3562A981 09...
Страница 228: ...HIOKI BT3562A981 09...