3.3 Common Operations
76
When measuring using the trigger functions, the timing of starting measurement
is different than that of starting recording (data acquisition).
See
"Measurement and Internal Operations" (p. 77)
To avoid operating mistakes when starting measurement, the behavior of the
START
key can be modified. At factory shipping, the
START
key is set to start
measurement when it is pressed once.
See
"13.2.1 Specifying Activation Conditions for the START Key" (p. 353)
While measuring, event marks can be inserted in recorded waveforms.
Event marks can then be searched after measurement.
See
"8.15 Inserting and Searching for Event Marks on a Waveform" (p. 231)
Press once
: recording stops at the end of the specified recording length.
Press twice
: recording stops immediately. (Abort)
When Aborting
•
Aborting while auto-saving
Data up to the moment of aborting is automatically saved.
•
Aborting while awaiting a trigger
If at least one trigger event has occurred since starting, the last measured
waveform is displayed. However, if longer than one half of the maximum setta-
ble recording length, no waveform is displayed.
•
Aborting while storing
Waveforms up to the moment of aborting are displayed.
An instrument setting can be changed to force measurements to be stopped by a
single press of the STOP key.
See
"13.2.2 Setting the Method for Stopping Measurement with the STOP Key" (p. 354)
3.3.6 Starting and Stopping Measurement
Starting Measurement
Press the
START/MARK
key.
The green LED at the left lights.
Start Measurement
Stopping Measurement
Press the
STOP
key.
Stop Measurement
Содержание 8860-50
Страница 2: ......
Страница 12: ...Contents x...
Страница 18: ...Reading this Manual 6...
Страница 24: ...1 3 Interconnection and Block Diagrams 12...
Страница 180: ...6 12 Making Trigger Settings on the Waveform Screen 168...
Страница 246: ...8 15 Inserting and Searching for Event Marks on a Waveform 234...
Страница 354: ...12 7 Print Examples 342 Print Example 4 Row Printing 1 4 steps Settings data Each step is 25 div...
Страница 356: ...12 7 Print Examples 344 A B Waveform_________________________________________________...
Страница 360: ...12 7 Print Examples 348 Event ________________________________________________________ Event mark...
Страница 406: ...14 2 External I O 394...
Страница 489: ......
Страница 490: ......