3.3 Common Operations
77
7
3
Chapter 3
Operatio
n Ov
erview
Measurement and Internal Operations _____________________________
Measurement methods are normal measurement (start recording when mea-
surement starts) and trigger measurement (start recording when trigger criteria
are satisfied).
In this manual, “Measurement start” means the instant when you press the
START
key, and “Recording start” means the instant when recording begins on
the waveform screen.
Trigger settings: "Chapter 6 Trigger Settings" (p. 135)
• Select the Trigger mode to record upon either single or repeating trigger
events. (p. 138)
• Enable pre-triggering if you want to capture data measured prior to trigger
events. (p. 140)
Start Measurement
START
key
End of Measurement
Recording
Starts
Recording
Stops
[Storing]
[Storing Done]
[ ]: Status Bar Display
Recording
Normal
Measurement
Without triggering
Stop Measurement
[Trigger Wait]
Recording
Starts
Recording
Stops
[Storing]
Recording starts when a trigger event occurs and continues for the specified recording
length.
Start Measurement
START
key
[Storing Done]
Recording
Trigger
Measurement
Trigger mode:
[Single]
Pre-triggering not
enabled
Single triggering
T
Recording starts when a trigger event occurs, continues for the specified recording
length, and returns to the Trigger Wait state.
Stop Measurement
STOP key
[Trigger Wait]
Recording
Starts
Recording
Stops
[Storing]
Start Measurement
START
key
Recording
[Trigger Wait]
T
Recording
Starts
Repeated triggering
Trigger mode:
[Repeat]
Pre-triggering not
enabled
T
[Pre-Trig Wait]
Stop Measurement
[Trigger Wait]
Start Measurement
START
key
T
Recording
Stops
[Storing]
After starting measurement and internally acquiring data for the
specified pre-trigger period, the
Trigger Wait state is activated.
Recording
Starts
The specified pre-trigger wait period is recorded before each trigger event
The data before a trigger event (for the pre-trigger period) is recorded.
Recording
[Storing Done]
T
Post-trigger waveform
Waveform during
specified pre-trigger
period
Repeated triggering and recording of phenomena before each event
Trigger mode:
[Repeat]
Pre-triggering enabled
Содержание 8860-50
Страница 2: ......
Страница 12: ...Contents x...
Страница 18: ...Reading this Manual 6...
Страница 24: ...1 3 Interconnection and Block Diagrams 12...
Страница 180: ...6 12 Making Trigger Settings on the Waveform Screen 168...
Страница 246: ...8 15 Inserting and Searching for Event Marks on a Waveform 234...
Страница 354: ...12 7 Print Examples 342 Print Example 4 Row Printing 1 4 steps Settings data Each step is 25 div...
Страница 356: ...12 7 Print Examples 344 A B Waveform_________________________________________________...
Страница 360: ...12 7 Print Examples 348 Event ________________________________________________________ Event mark...
Страница 406: ...14 2 External I O 394...
Страница 489: ......