375
14
Chapter 14
External Control
This section explains the terminals used for external control of the instrument.
As shown below, there are two types of terminals. The general term “external
control terminal” refers to both types.
External Control
Chapter 14
Right-side panel
EXT.SMPL terminal
SYNC.OUT terminal
Input
Output
Make signal setting on the Ext Term
(External terminal) Settings screen.
External control terminal
External input (p. 391)
Control measurement start and stop, printing, and
saving with external signals.
Output GO/NG calculation evaluation
results (p. 387)/
External output (p. 389)
Output a signal when the evaluation results of nu-
merical calculation are GO or NG. Also used to
signals for specified instrument states, such as
BUSY, saving data, error, and so on.
Trigger output (p. 380)/
Probe calibration signal output (p. 393)
Output a signal when a trigger event occurs. Also
used to calibrate attenuating probes before using
them.
External trigger input (p. 378)
Input external signals.
GND (common with instrument ground)
Sample synchronizing signal output
(p. 384)
Use to synchronize sampling with multiple instru-
ments.
External sampling (p. 382)
Input an external sampling signal. (Memory func-
tion and FFT function only)
Содержание 8860-50
Страница 2: ......
Страница 12: ...Contents x...
Страница 18: ...Reading this Manual 6...
Страница 24: ...1 3 Interconnection and Block Diagrams 12...
Страница 180: ...6 12 Making Trigger Settings on the Waveform Screen 168...
Страница 246: ...8 15 Inserting and Searching for Event Marks on a Waveform 234...
Страница 354: ...12 7 Print Examples 342 Print Example 4 Row Printing 1 4 steps Settings data Each step is 25 div...
Страница 356: ...12 7 Print Examples 344 A B Waveform_________________________________________________...
Страница 360: ...12 7 Print Examples 348 Event ________________________________________________________ Event mark...
Страница 406: ...14 2 External I O 394...
Страница 489: ......
Страница 490: ......