5.4 Converting Input Values (Scaling Function)
123
5
Chapte
r 5
Input Chan
nel Setting
s
About the Scaling
Function
Use the scaling function to convert the measured voltage units output from a
sensor to the physical units of the parameter being measurement.
Hereafter, “scaling” refers to the process of numerical value conversion using the
Scaling function.
Gauge scales, scale values (upper and lower limits of the vertical axis) and A/B
cursor measurement values can be displayed in scaled units. Scaling is avail-
able for each channel.
Scaling Setting Example
See
When using a clamp sensor (p. 126) (Example: Converting [ V ]
[ A ] )
When using the Strain Unit (p. 127) (Example: Converting [
]
[ G ])
Scaling Methods
Two scaling methods are available:
5.4
Converting Input Values (Scaling Function)
When scaling is enabled, the space between the brackets [ ] is widened.
Before Scaling
After Scaling
Set the physical value per volt (conversion
ratio: eu/V) of the input signal, an offset value
and measurement unit name (eu: engineering
units) for conversion, so measurement values
acquired as voltage are converted to the speci-
fied units.
Example:
Conversion ratio: A value per volt,
Offset value: B, Unit name: A
Conversion Ratio Setting
[A]
[V]
Convert from slope (conversion
ratio) and offset value
B
(Example: Converting [V]
[A])
Set the voltage values of two points of the input
signal, the converted unit value of these two
points and the name of the converted measure-
ment units, so measurement values acquired as
voltage are converted to the specified units.
Example:
Unit name: A
Voltage value at 2 points Voltage of units to convert
V
H
: Higher potential point A
H
:Value for higher
potential point
V
L
: Lower potential point A
L
: Value for lower
potential point
Two-Point Setting
A
H
A
L
V
L
V
H
[A]
[V]
Conversion ratio and offset val-
ue are calculated from the two
points and converted
Actual measurement
values
Converted unit values
Содержание 8860-50
Страница 2: ......
Страница 12: ...Contents x...
Страница 18: ...Reading this Manual 6...
Страница 24: ...1 3 Interconnection and Block Diagrams 12...
Страница 180: ...6 12 Making Trigger Settings on the Waveform Screen 168...
Страница 246: ...8 15 Inserting and Searching for Event Marks on a Waveform 234...
Страница 354: ...12 7 Print Examples 342 Print Example 4 Row Printing 1 4 steps Settings data Each step is 25 div...
Страница 356: ...12 7 Print Examples 344 A B Waveform_________________________________________________...
Страница 360: ...12 7 Print Examples 348 Event ________________________________________________________ Event mark...
Страница 406: ...14 2 External I O 394...
Страница 489: ......
Страница 490: ......