8.9 Message Reference
234
Setting and Query of Measurement Value Memory Function Memory Size
Syntax
Command
Query
Response
:MEMory:POINts
<Numeric Value>
:MEMory:POINts?
<Numeric Value>
= 1 to 32,000 (NR1)
Explanation
Command Sets the memory size (the number of measurements to save).When
the setting is changed all saved measurement values are erased.
A numeric value in NRf format is accepted but non significant digits
are rounded off so the numeric.
Query
Returns the memory size as an NR1 numerical value.
Example
Command
Query
Response
:MEMory:POINts 200
Sets the memory size to 200.
:MEMory:POINts?
:MEMORY:POINTS 200
(when HEADER ON)
200
(when HEADER OFF)
The memory size is set to 200.
Query of Voltage and Current Monitor Values
Syntax
Query
Response
:MONItor?
<Voltage Monitor Value(NR3)>,<Current Monitor Value(NR3)>
Explanation
Query
Returns the measurement signal monitor values in the order of volt-
age monitor value and current monitor value.
Example
Query
Response
:MONItor?
9.56789E-01,7.34567E-05
The voltage monitor value and current monitor value are set to
0.956789 V and 73.4567
A.
Содержание 3504-40 C HiTester
Страница 2: ......
Страница 22: ...1 4 Names and Functions of Parts 16 ...
Страница 28: ...2 5 Turning the Power On and Off 22 ...
Страница 40: ...3 3 Setting the Measurement Conditions 34 OF UF Evaluation Flow Chart ...
Страница 66: ...4 4 Self Calibration 60 ...
Страница 96: ...5 2 BIN Measurement Function Model 3504 50 3504 60 only 90 ...
Страница 136: ...7 4 About Measurement Times 130 ...
Страница 279: ...Appendix 5 External View Appendix A7 Appendix 5 External View Unit mm 220 3 260 1 100 1 32 5 25 21 54 20 5 95 5 15 ...
Страница 286: ...Appendix 7 Initial Settings Table A14 ...
Страница 289: ......
Страница 290: ......
Страница 291: ......
Страница 292: ......