3.1 Adjusting and Correcting Values
_______________________________________________________
NOTE
NOTE
3.1.2 Zero-Cancel Correction Function
Use the zero-cancel correction function if the
counter fails to revert to zero after correct auto-zero-
adjustment in AC+DC A mode.
In a current mode, if you press
0ADJ/RESET
button
again during the auto-zero-adjustment in the internal
circuit, the auto-zero-adjustment is canceled.
The auto-zero-adjustment is not effective at auto
range in DCV mode. If you perform the auto-zero-
adjustment at manual range, the auto-zero-
adjustment is invalid by changing the range.
Use the zero-cancel correction function when the
counter fails to become zero under no input in AC
A, AC+DC A, AC V, or AC+DC V mode.
1. Press
HOLD
button to display
HOLD
annunciator.
2. Press
0ADJ/RESET
button.
ADJ
annunciator blinks.
When there is an input or the counter decreases, the
measured values will be evaluated lower by pressing
0AJD/RESET
button.
If the counter is zero, the zero-cancel correction
function does not work.
In AC+DC A mode, the zero-cancel correction
function does not work, unless auto-zero-adjustment
is complete.
The zero-cancel correction function is not effective
at auto range in DCV mode. If you perform the
zero-cancel correction at manual range, the zero-
cancel correction is invalid by changing the range.