Urgent Medical Device Correction - Ear Thermometer 

OVERHEATING ASSOCIATED WITH THE  

BRAUN THERMOSCAN PRO 600 EAR THERMOMETER.  

 
 

 

 

 

 

 

 

 

 

 
  
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 

SAFETY ALERT

 

 

Author:  

Toni Young, MSN, CVAHP 

Supply Chain Management 

 

Date: 

 08/06/2021 

 

HRO PRINCIPALS USED:  

 

Pay Attention to Detail:  

when cleaning and using the Braun ThermoScan Pro 600 Thermometer.

 

 

Stop – Validate – Verify

 

before using the

 

thermometer that the

 

thermometer is functioning properly and does 

not have a blinking green light, require multiple restarts, or if the tip overheats.

 

 

NEXT STEPS

:

  

Please 

share this Safety 

Alert at your Safety 

Huddles and 

Department 

meetings. 

 

Please post and 

follow proper 

cleaning procedure, 

attached.

 

 

SITUATION:   

There  is a potential risk associated  with overheating the 

Braun ThermoScan PRO 6000 probe tip due to fluid ingress. If the device 
is  exposed  to  fluid  ingress  and  is  used  before  the  cleaning  fluid  has 
completely  dried  there  is  a  risk  of  the  device  overheating  potentially 
causing a burn to the user or to the ear canal of the patient.  

 

 
BACKGROUND

:  

Hillrom has received 116 reports of complaints regarding 

overheating of the Braun ThermoScan 6000 probe tip out of more than 
1.1 million units sold.  To date, they are aware  of ONE reported  injury 
associated with this issue. 

 

 
ASSESSMENT:

  

1.

 

Use the correct cleaning procedure for the device.  

 

2.

 

Inform Users of the potential safety risk of inappropriate cleaning

 

3.

 

Inform users of what  visual signals  to look for to  minimize  the potential to 
overheating probe tips.

 

RECOMMENDATION:

  

1.

 

Share this communication with all potential users and provide the cleaning 
guide for proper cleaning.   

2.

 

Contact Supply Chain if the following occurs: 

a.

 

Do not  use the device if the ring around the measurement button 
shows  a  green  blinking  or  flashing  light  instead  of  a  “ready  state” 
(solid green light).  

b.

 

Do not use the device if the device requires multiple power-ups prior 
to going to the “ready state” (solid green light). 

c.

 

If you experience an overheating probe tip, do not use the device. 

 