General Cleaning/Disinfecting
35
Disinfecting
When there is visible soilage and between patients, we recommend that you disinfect the unit with a
tuberculocidal disinfectant. (For customers in the US, the disinfectant should be registered with the
Environmental Protection Agency.)
Dilute the disinfectant according to the manufacturer’s instructions.
Cleaning Sleep Surface
Follow these instructions for cleaning the Sleep Surface:
•
Unplug the system from its power source.
•
To clean directly beneath the sleeping surface at the head end, lift the head end of the mattress.
•
To clean directly beneath the sleeping surface at the foot end, lift the foot end of the mattress.
To remove the sleep surface mattress:
•
Lift up on the foot and head ends of the sleep surface until the magnets inside the ticking release the frame.
Slide the sleep surface off the articulating deck.
To replace the sleep surface mattress, do the following:
•
Return the patient deck to the level position, and fully extend the foot section.
•
Position the sleep surface on the system frame.
•
Ensure that the foot section of the sleep surface is extended completely to the foot board.
Cleaning Medical Fluid Spills
Fluid spills should be wiped up as soon as possible. Always unplug the unit from its power source before cleaning up
major fluid spills. Some fluids used in the hospital environment, such as iodophor and zinc oxide creams, can cause
permanent stains.
Temporary stains can be removed by wiping vigorously with a lightly-dampened sponge or rag and an approved
cleaner/disinfectant solution.
Cleaning Blood And Excreta
If possible, wipe up excess blood and excreta when wet since the cleaning process is more difficult after these
substances dry on the surface. In the presence of visible blood or other body fluids, the use of an intermediate-level
(tuberculocidal) detergent/disinfectant would be recommended.