4.37 Stow-Away Push Handle (Stretchers with the Integrated Oxygen Tank Storage)
Chapter 4: Removal, Replacement, and Adjustment Procedures
4.37 Stow-Away Push Handle (Stretchers with the Integrated
Oxygen Tank Storage)
Tools required:
T27 Torx®
1
screwdriver
Retaining ring removal and installation tool
Removal
1. Make sure the brakes are set.
2. Raise the head section to its highest position.
3. Remove the utility tray (A) (see figure 4-60 on page 4-116).
Figure 4-60. Removal
(Note: The push handle shown includes the optional integrated IV pole handle.)
4. Remove the retaining ring (B) and bumper (C) from the handle (D).
5. Loosen the two screws (E) that hold the handle (D) in the adapter (F).
6. Remove the handle (D).
1. Torx® is a registered trademark of Acument Intellectual Properties, LLC.