
44
Hill-Rom® P500 Therapy Surface User Manual (158251 REV 3)
Deflate Procedure
CAUTION:
The Service mode is intended for use by authorized persons. Get
access to the Mattress Deflate menu only. Otherwise, equipment damage or fail-
ure could occur.
Make sure the hose is connected to the control unit and the surface.
Select
Yes
.
Select
Mattress Deflate
.
Select
Configurations
.
The display shows:
Rental
Capital
NOTES:
1.
The
(+)
and
(-)
controls move you through the menu items. The
lb/kg
con-
trol selects an item.
2.
Deflation takes approximately 5 minutes for a Rental unit and 10 minutes
for a Capital unit.
3. Capital units
—as the display toggles between Mattress Venting and Exit
when finished, push down on the surface to help it deflate more quickly.
Содержание P005723
Страница 1: ...USER MANUAL Hill Rom P500 Therapy Surface Product No P005723 and P005787 158251 REV 3...
Страница 2: ......
Страница 4: ...ii Hill Rom P500 Therapy Surface User Manual 158251 REV 3...
Страница 53: ......