Assembly

Check the product before each use.

 Check the following points with regard to wear and damage:

•  Straps
•  Seams

 

Do not use damaged lifting accessories.

  

If anything is unclear, please contact the manufacturer or supplier.

Care and Maintenance

Washing instructions: 

Please see the Extension Loop product label and the attached appendix ”Care and Maintenance of Liko 

Slings

.

Periodic Inspection

The product must be inspected at least once every 6 months. More frequent inspections might be required 

if the products 

are used or washed more often than normal. 

For protocols please contact your Hill-Rom representative.

Expected Life Time

If the product is made of green woven polyester webbing, the product has an expected life time of 5 years. If the product is 

made of grey polyester, the product has an expected life time of 3 years. The expected life time is based on 10 uses per day 

and washing once per week, according to washing instructions. Hill-Rom recommends the product to be used only during its 

expected life time.

 

Recycling Instructions

The entire Extension Loop should be sorted as Combustible waste. 
Hill-Rom evaluates and provides guidance to its users on the safe handling and disposal of its devices to aid in the prevention 

of injury, including, but not limited to: cuts, punctures of the skin, abrasions, and any required cleaning and disinfection of the 

medical device after use and prior to its disposal. Customers should adhere to all federal, state, regional, and/or local laws and 

regulations as it pertains to the safe disposal of medical devices and accessories. 
If in doubt, the user of the device shall first contact Hill-Rom Technical Support for guidance on safe disposal protocols.