![Hill-Rom Liko Viking S Скачать руководство пользователя страница 12](http://html1.mh-extra.com/html/hill-rom/liko-viking-s/liko-viking-s_instruction-manual_2134184012.webp)
© Copyright Liko
AB 201
1-05
Inspection and Maintenance
Care and Maintenance
For trouble-free use, certain details should be checked each day the lift is used:
• Inspect the lift and check to make sure that there is no external damage.
• Check the sling bar attachment.
• Check the functionality of the latches.
• Check the integrity of the lifting motion and the base-width adjustment.
• Check to make sure that the emergency lowering (both electrical and mechanical) works.
• Charge the batteries each day the lift is used and make sure the charger works.
When necessary, clean the lift with a moist cloth, using warm water or disinfectant and check that the wheels are
free from dirt.
NOTE! Do not use cleaning agents that contain phenol or chlorine, since these can damage
aluminium and polyamide material.
The lift should not be exposed to running water.
Service
Viking S and Viking XS
should be periodically inspected at least once a year.
Periodic inspection, repair and maintenance should be performed only in accordance with the Liko
Service Manual and by personnel authorized by Liko and using original Liko spare parts.
Service Agreement
Liko offers the opportunity to enter into service contracts for the maintenance and regular inspection of your Liko
product.
Expected Life Time
The product has an expected life time of 10 years when correctly handled, serviced and inspected in accordance
with Liko’s instructions.
Transport and Storage
During transportation, or when the lift is not to be used for a long time, the emergency stops should be engaged.
The environment where the lift is transported and stored should have a temperature of 10–40 °C (50–104 °F)
and a relative humidity of 30–75 %. The air pressure should be 700–1060 hPa.
Recycling
For instructions on how to recycle your Liko product, please visit our website: www.liko.com.
Product Changes
Liko’s products undergo continuous development, which is why we reserve the right to make product changes
without prior notice. Contact your Liko/Hill-Rom representative for advice and information about product upgrades.
Design and Quality by Liko in Sweden
Liko is quality-certified in accordance with ISO 9001 and its equivalent for the medical device industry, ISO 13485.
Liko is also certified in accordance with environmental standard ISO 14001.
Manufacturer:
Liko AB
SE-975 92 Luleå
Sweden
www.liko.com