4.2 HI/LO Actuator
Chapter 4: Removal, Replacement, and Adjustment Procedures
Page 4 - 6
Electrically Operated General Ward Bed
Operation and Maintenance Manual
(154588 REV 1)
8. Disconnect the cable-to-control box connection as required to allow the
actuator to be removed from the bed.
NOTE:
The type of cable from the actuator will vary depending on the configuration
and date of manufacture.
9. Remove the retaining clip, and unplug the in-line socket (see figure 4-6 on
Figure 4-6. Mini-Fit Cable Configuration
– For hard-wired configurations, access the sockets by removing
the saddle. Use a flat screwdriver to pry in a downward direction
(see figure 4-7 on page 4-6).
Figure 4-7. Hard-Wired Cable Configuration