Service Manual LikoGuard™
overhead lift
• 3EN190401 Rev. 7
24
www.hillrom.com
7.0 Periodic Inspection Information, LikoGuard™ overhead lift
According to ISO 10535:2006 Annex B, Periodic Inspection of medical device Liko™ overhead lift
should be performed at least once a year. For LikoGuard™ overhead lift Periodic Inspection overhead lift
(3EN191001) is available on the Partner Web and in Agile (internally).
Liko™ overhead lifts must be thoroughly inspected at least once per year. Inspection and service must be carried out by Hill-Rom authorized personnel.
If the lift is installed in a corrosive environment such as indoor pool or bathroom, see inspection point 11 before starting inspection.
NOTE! For inspection of the rail system, use Periodic Inspection Liko™ Rail System (3EN111001)
If printed: Make a colour print of this instruction.
Periodic Inspection
Liko™ overhead lifts
3EN191001 Rev. 2
Valid for following overhead lifts: LikoGuard™, Likorall™, Multirall™, Masterlift™
Lift information
Customer reference
Lift:
Agreement No:
Product Number:
Name:
Serial Number:
Address:
Production year:
Post Code
Inspection sign off
Complete inspection according to Instructions for the check points (page 2-8), fill in this page and sign below.
Approval to use the overhead lift
Approved
Not approved
If the overhead lift has one or more inspection points “Not approved”, the lift must not be used.
Action required:
Actions according to the inspection items "NOT APPROVED" should be performed immediately. After performed actions
sign below. If anything is unclear or if you have questions, please contact Hill-Rom or your local Hill-Rom representative.
Contact information is to be found at www.hill-rom.com.
Inspection performed by:
Date:
Next inspection:
Inspection performed in accordance with ISO 10535:2006 Annex B- Periodic inspection
OVERHEAD LIFT
Action required:
Approved
Not
approved
1
Product Decal
2
General Inspection
3
Carriages
4
Hand Control
5
Emergency Stop
6
Emergency Lowering
7
Limit Switch
(Not valid for MasterLift)
8
Charger
(Not valid for MasterLift)
9
SlingBar
10 Lift Strap
ENVIRONMENTAL IMPACT N/A
Action required:
Approved
Not
approved
11 Corrosive environments
LOAD TESTING
Max Load applied:
Approved
Not
approved
12 Maximum Load overhead lift
Kg: Lbs:
Emergency lowering Maximum Load
Kg: Lbs:
Required Measurements:
Measured dimensions:
Approved
Not
approved
12 Load testing / Lift strap drift
mm: Inch:
Clear service
N/A
Action required:
Approved
Not
approved
13 Service Symbol reset
Liko technical documentation can be found on the Partnerweb:
https://partner.liko.com - “My information” - “Documents”
If you don’t have access, send a mail for registration to: [email protected]
Periodic Inspection may only be performed by personnel authorized by Hill-Rom and using
original Liko™ spare parts.
Available Liko Periodic Inspections:
•
3EN371001 Periodic Inspection Liko™ Mobile lifts
•
3EN111001 Periodic Inspection Liko™ OH Rail system
•
3EN191001 Periodic Inspection Overhead Lifts
• 3EN301001 PI Freespan Freestand
•
3EN601001 Periodic Inspection Liko™ accessories
Also available is Periodic Inspection Slings. This inspection should be performed by a person who is
suitable, has good experience and sufficient knowledge about design, use and maintenance of the sling.
•
7EN160834 General Periodic Inspection Protocol Liko Slings and Sabina SupportVests
NOTE!
Periodic Inspection shall be performed at least once per year.
INF
ORMA
TION