Safety information
Warning of premature pro-
gram abort
Warning of hand injury due to
rotor gears
Risk of injury due to the
device falling out of the pack-
aging
Warning of hand injury due to
device lid falling closed
unbraked
WARNING
Warning of premature program abort!
In the event of a premature program abort, such as a power
failure, switching off during the program run or pulling out
the mains plug, the desired effect on the samples might not
be achieved.
−
Do not switch off the device during a program run.
−
Do not actuate the emergency release during a program
run.
−
Do not pull out the mains plug during a program run.
−
After a program abort, check whether the samples can
be used.
CAUTION
Warning of hand injury due to rotor gears!
There is a risk of injury to fingers by the rotor gears if this is
turning. Cuts with possible infections could result.
−
Do not grab between the gears of the rotating units and
the rotor.
−
Avoid rotor rotation with the lid open as much as pos-
sible.
−
To prevent injuries to third parties, do not leave the
device unsupervised with the lid open.
WARNING
Risk of injury due to parts falling out of the packaging!
When parts fall out of the transport packaging, there is a risk
of crushing injuries to extremities.
−
Observe the warnings for opening the packaging.
−
Only open the packaging at the specified point.
WARNING
Warning of hand injury due to device lid falling closed
unbraked!
If the lid falls closed unbraked, there is a risk of extremities
getting caught.
−
Have the device maintained by a service technician at
the specified intervals.
−
Always open the device lid completely.
−
If a gas pressure spring of the lid is defective, have it
replaced by a service technician.