161 / 174
en
12.2020 / Rev. 02
AB66005
Examples
Fig. 222: Temperature drops
1.
In manual mode, set the temperature drop to 4°C from Friday
afternoon at 2:00 p.m. to Monday morning at 3:00 a.m. and
set the repetition interval to
‘weekly’
2.
Set another temperature drop from Tuesday evening at 20:00
p.m. to Thursday morning at 03:00 a.m. without a repetition.
3.
Set the target temperature to 37°C
4.
Start manual mode
Chapter 6.11.2 ‘Starting manual mode’
B.4
Temperature drop on the weekend with start delay
Mon
Tue
Thu
Fri
Sat
Sun
Mon
Wed
37
°C
4
1
Fig. 223: Diagram ‘Temperature drop on the weekend with start
delay’
1 Start delay
Operating mode:
Manual mode; Start delay and temperature drop (both are activated
at the same time)
Procedure:
1.
In manual mode, set the temperature drop to 4°C from Friday
afternoon at 2:00 p.m. to Monday morning at 3:00 a.m. and
set the repetition interval to
‘weekly’
2.
Set the target temperature to 37°C
3.
Set the starting date in the start delay
Until manual mode is started, the incubator
remains unregulated.