Working Rules 

 

Rules to follow prior to use... 

 

-  The cabinets should be used indoors in a normal clinical environment with a temperature between 

18 

°

C and 28 

°

C, normal humidity and atmospheric pressure. 

 
-  All the cabinets are designed and produced to be used in clean environments. 
 
-  They are not to be used outdoors or in environments with high humidity or pollution. The cabinets 

are not constructed to ventilate vapours with acid or containing organic solvents. The cabinets 
must not be used as fume hoods. 

 
- About 10 minutes prior to usage, the fan of the unit should be switched on for operation at normal 

speed. 

 

- The work chamber and front/side windows are to be carefully cleaned and disinfected. Use 70 % 

ethanol or similar. It is recommended to use special lint-free wipe. 

  
- Objects and appliances must be carefully cleaned or disinfected prior to being introduced into the 

work chamber. 

 
- The front window is positioned in working position and kept in that position during the 
  entire work process. 
 
- Necessary appliances for use during work must be placed within easy reach. 
 
- Ensure appropriate protection of operator as well as product (e.g. clothes, gloves, masks, visors,  
  etc.). 
 

Important when performing biohazard work!

 

 

- Never perform any work with the fan at reduced speed. 
 
- The front window must be in working position during the work.  
 
- Place the product behind the front suction holes of the table top,- as far back as possible. 
 
- Work with steady/soft movements. 
 
- Never overload the work chamber. 
 
- Reduce the number of transfers into and out of the work chamber. 
 
- Avoid products or remedies with strong emission of heat. 
 
- Do not position the cabinet in places with direct draught towards the work opening. 
 
-Position the cabinet so that frequent passing of persons in front of the work opening is avoided.

 

F

 

Содержание 47240000

Страница 1: ...Erreur Argument de commutateur inconnu INSTRUCTION MANUAL HOLTEN LAMINAIR SAFE 2000 MAXI SAFE 2000 ID 807222...

Страница 2: ...r Argument de commutateur inconnu 15INSTRUCTION MANUAL SAFE 2000 MAXI SAFE 2000 Erreur Argument de commutateur inconnu Fig A Control Panel Fig B LCD Display 1 2 3 4 5 6 7 8 9 10 11 12 13 14 15 16 17 1...

Страница 3: ...inconnu 15 Erreur Argument de commutateur inconnu Contents Page 1 Introduction 3 2 Safety Precautions 3 3 Installation 3 4 Working Rules 3 5 Operation 3 6 Technical Specifications 3 7 Maintenance 3 E...

Страница 4: ...pressure plenum for highest operator and product safety Motor driven double function front window with both sliding and hinged modes Key switch for pre selection of desired fan speeds to be eligible...

Страница 5: ...ree access to the intended site of installation The cabinet may be tilted 10 over an axle so that the sloping front is in vertical position Furthermore you may dismount the front shield with the contr...

Страница 6: ...to an installation which is protected by high sensitive leakage current contact breaker 6 Valves for gases vacuum or liquids are positioned in a separate plate in the left or right hand side window an...

Страница 7: ...recommended to use special lint free wipe Objects and appliances must be carefully cleaned or disinfected prior to being introduced into the work chamber The front window is positioned in working posi...

Страница 8: ...utton to confirm a selection during settings 9 Key switch for activation deactivation of the start stop button and switching between reduced or normal fan speed 10 Button to lower the front window Fig...

Страница 9: ...tion the UV icon will appear in the display 2 Press the ENTER button and the display will flash for setting of minutes 3 Press the button to set the number of minutes select from 1 to 59 minutes 4 Pre...

Страница 10: ...the lock switch a Lock switch in position 1 m Press the FAN button Once to select normal speed of the fan Twice to select reduced speed of the fan Three times to select the fan off b At priority Turn...

Страница 11: ...SAFE 2000 MAXI SAFE 2000 INSTRUCTION MANUALPage Erreur Argument de commutateur inconnu 15 Erreur Argument de commutateur inconnu...

Страница 12: ...22 0 l 7 5 l 27 0 l 9 0 l Exhaust air volume 300m3 h 10 400m3 h 10 500m3 h 10 600m3 h 10 Power 2 A 2 A 3 5 A 3 5 A Power consumption 0 5 kW 0 5 kW 0 8 kW 0 8 kW HEPA filter efficiency 99 999 99 999 9...

Страница 13: ...operation of the safety cabinet and compliance with standards are based on the following conditions 1 Correct air velocities 2 Efficiency of the installed HEPA filters 3 Tightness of the construction...

Страница 14: ...Page Erreur Argument de commutateur inconnu 15INSTRUCTION MANUAL SAFE 2000 MAXI SAFE 2000 Erreur Argument de commutateur inconnu...

Страница 15: ...SAFE 2000 MAXI SAFE 2000 INSTRUCTION MANUALPage Erreur Argument de commutateur inconnu 15 Erreur Argument de commutateur inconnu Manual 00346609 replacing 00346608 990104...

Страница 16: ...chlorine dioxide hypochloride phenol and perchloric acid Useable disinfectants contain alcohol or aldehydes Recommended disinfectants are 70 volume pure ethanol or 2 volume glutoric aldehyde or 15 vo...

Страница 17: ...BOOK Type Product name Serial no Supplier Service interval Laboratory Service repair Fault occurred on Notified date Repaired date Remark fault Caused by action Service internal external Carried out...

Страница 18: ...Service repair Fault occurred on Notified date Repaired date Remark fault Caused by action Service internal external Carried out by...

Страница 19: ...Service repair Fault occurred on Notified date Repaired date Remark fault Caused by action Service internal external Carried out by...

Страница 20: ......

Страница 21: ...of machinery Basic concepts General principles EN 292 2 1993 Safety of machinery Technical principles and specifications EN 60204 1 1992 Safety of machinery Electrical equipment of machines General re...