3. FAULTS – CAUSES AND REMEDIES
DISPLAY
CAUSES
REMEDY
Underload: only dashes appear at the
bottom of the display field.
Switch the display off and on. The
zero point is reset automatically.
Overload: only dashes appear at the top of
the display field. Maximum weighing range
is exceeded.
Unload the scales, or switch them
off and on again. The zero point is
reset automatically.
Scales were not stabilised when switched
on.
Check the surface under the
scales, and check that the weighing
platform is level. Test the scales
with a force shunt. If the error
persists after readjustment, please
contact your service partner.
Taring not possible.
Stand still, or stabilise the weight
on the scales.
If the fault cannot be cleared, or other error messages appear, please contact your