Precision Balance
- 6 -
Linearity Calibration of Weight
Step I
Power off while holding down
keys, and then press
「
ON/OFF
」
to switch on the scale.
Step II
Clear the weighing pan, then press
key, (zero point
calibration)
Step III Wait for a beep sound
Step IV
Place 1/3 of load on the weighing pan and press
to
perform 1/3 load calibration. Wait for a beep sound.
Step V
Place 2/3 of load on the weighing pan and press
to
perform 2/3 load calibration. Wait for a beep sound.
Step VI
Place full load on the weighing pan and press
to perform
full load calibration. Wait for a beep sound.
Step VII
Clear the weighing pan, then press
key
Step
VIII
After countdown, the linearity calibration is completed and the
scale returns to the normal weighing mode.
Содержание HXB series
Страница 1: ...PRECISION BALANCE USER S MANUAL MODEL X 2007 V2 0 ...