76
Known Interference
Nucleated red blood cells (NRBC) may be counted as white
blood cells and give falsely elevated total leukocyte count. This
will mainly be reflected in the differential count as an elevated
lymphocyte count.
Cold agglutination or cryoglobulin may interfere.
Low Hb (<80 g/L) may cause an increased level of flags.
NOTE: Discard sample if clots.
Technical Specifications
Dimensions: 188x157x155 mm (7.40 x 6.18 x 6.10 inches)
Weight: 1300 g (2.87 pounds) (with 6 C (LR14/HR14) batteries
installed)
AC adapter: CE marked
Only use AC adapters listed under AC Adapters.
Pollution degree: 2
Overvoltage category: II
Atmospheric pressure: 700 hPa to 1060 hPa.
Equipment not suitable for use in the presence of flammable
mixtures.
The HemoCue WBC DIFF system has been tested for electrical
safety and EMC according to the following standards:
• IEC 61010-1 Safety requirements for electrical equipment for
measurement, control, and laboratory use - Part 1: General
requirements
• IEC 61010-2-101 Safety requirements for electrical equipment
for measurement, control and laboratory use – Part 2-101:
Particular requirements for in vitro diagnostic (IVD) medical
equipment.
• UL 61010/CSA-C22.2 No. 61010-1 Safety requirements for
electrical equipment for measurement, control and laboratory
use – Part 1: General requirements.
• IEC/EN 60601-1-2 Ed 3.0 Medical electrical equipment-Part 1:
General requirements for Safety – Part 1-2: Collateral Standard:
Electromagnetic compability – requirements and tests.
Essential Performance
The HemoCue WBC DIFF system is an
In-Vitro
diagnostic system
designed for quantitative determination of white blood cells
(WBC) in capillary or venous whole blood. The system provides
values for a total white blood cell count and a differential white
blood cell count including neutrophil count, lymphocyte count,
monocyte count, eosinophil count and basophil count.
Содержание WBC DIFF
Страница 1: ...HemoCue WBC DIFF Operating Manual ...
Страница 4: ...3 2 1 5a 7 4 6 5b 4 Components ...
Страница 6: ...1 2 3 6 Start Up ...
Страница 8: ...5 8 6 7 4 8 ...
Страница 10: ...1 2 3 10 Data Entry Navigation ...
Страница 18: ...1 2 3 4 5a 5b 18 Set Up Settings ...
Страница 20: ...6 7 8 9 10 11 20 ...
Страница 22: ...12 13 14 15 16 22 ...
Страница 24: ...17 18 19 20 21 24 ...
Страница 26: ...1 2a 2b 26 Set Up Printer ...
Страница 28: ...1 28 Set Up Keyboard ...
Страница 30: ...1 30 Set Up Barcode Reader ...
Страница 32: ...1 32 Set Up Connectivity 2 ...
Страница 34: ...1 3 2 34 Blood Collection Capillary Sample ...
Страница 36: ...4 5 6 7 36 ...
Страница 38: ...8 9 10 38 ...
Страница 40: ...1 2a 2b 3 4 5 40 Blood Collection Venous Sample and Control Material ...
Страница 42: ...1 2 4 3 5 42 Routine Use Patient Test ...
Страница 44: ...6 7a 7b 8 44 ...
Страница 46: ...10 9a 9b 9c 46 ...
Страница 48: ...1 2 4 3 6 5 48 Routine Use QC Test ...
Страница 50: ...50 7 8 9a 9b 10 ...
Страница 52: ...12 11a 11b 11c 52 ...
Страница 54: ...1 2 3 4 5 6 54 Review Delete Results ...
Страница 56: ...7 8 9 56 ...
Страница 58: ...1 2 3 4 5 58 Maintenance Daily Maintenance ...
Страница 60: ...60 1 2b 3 Maintenance Optical Parts 2a 4 ...