144
FIGURE 6-25
When all levels have been completed, an
image containing all results is displayed.
If the mean result (29) for each level is within
the concentration range (30) for the level,
the Qualitative test result (31) will indicate
“Within range”.
If the mean result (29) for any level is beyond
the concentration range for the level (30),
the result and the range limits will appear in
bold, and the Qualitative test result (31) will
indicate “Out of range”.
w) Press the Confirm button (32).
27 Total No. of Cuvettes used in the linearity test
28 Unit for analyte concentration
29 Mean result at the level
30 Concentration min. and max. value at the level
31 Qualitative Test Result
32 Confirm button
FIGURE 6-26
If no ranges are specified for the
concentration levels (see
the mean results at each level (34) are listed
and no range limits will be displayed. No
Qualitative test result is given.
a) Press the Confirm button (35).
33 Total No. of Cuvettes used in the linearity test
34 Unit for analyte concentration
35 Mean result at the level
36 Confirm button
Linearity
Result
N = 15
mg/dL
1 = 75 (70 - 80)
2 = 143 (140 - 150)
3 =
209 (210 - 220)
4 =
278 (280 - 290)
5 = 354 (350 - 360)
Out of Range
27
31
29
30
28
32
Linearity
Result
N = 15
mg/dL
1 = 75
2 = 143
3 =
209
4 =
278
5 = 354
35
34
33
36