1-360008/G 

18

 

to stabilize and then take a temperature reading from a calibrated thermometer inside the chamber 
bath.  If the thermometer reads the same as the controller display, the temperature controller is 
calibrated correctly.  If there is a temperature variance, the temperature controller needs to be 
calibrated. 
 
As an example, if a calibrated thermometer reads 35.5

º

C and the Temperature Controller reads 36.5

º

C, 

then the controller display ("CAL" function) needs to be reduced by 1.0

º

C.  If the thermometer reads 

37.5

º

C and the Temperature Controller reads 36.5

º

C, then the controller display ("CAL" function) 

needs to be increased by 1.0

º

C. 

 

To recalibrate the Temperature Controller: 

 

1)   Press and hold the Up and Down arrow keys until a controller program is displayed.  (If the 

"CnFg" prompt appears first, the Controller will need to be unlocked - refer to "Accessing the 
Program Lockout"). 

 

2)   Press the Down arrow key until the "PId" Menu prompt appears. 

 

3)   While pressing the SET key in (the digital readout will display "no"), press the Down arrow 

key to display "yes", and then release the SET key. 

 

4)   Press the Up arrow once to display the "CAL" prompt. 

 

5)  While holding the SET key down, press either the Up or Down arrow keys to alter the 

calibration program value. 

 

6)  To exit back to the main operating mode, release the set key and press the Up and Down 

arrow keys simultaneously until the operating temperature appears on the digital display. 

 
After making any calibration changes to the Temperature Controller allow the chamber temperature to 
stabilize and make a new reading to verify that the controller is properly calibrated.  Make any 
additional adjustments as needed until the controller readout is properly calibrated. 
 
 
 
 
 
 
 
 
 
 

 
B -Alarm System Test 

 
To test that the High Alarm System is working properly, it is recommended to follow methods that 
may be recommended by the AABB or other governing organizations for such alarm system checks.  
A suggested method to activate an actual high alarm test condition is as follows: 
 

Содержание QuickThaw DH4

Страница 1: ...NG SYSTEM MODELS DH4 and DH8 VERSION K Operation Service Maintenance HELMER LABS INC 15425 HERRIMAN BLVD NOBLESVILLE IN 46060 USA PHONE 317 773 9073 FAX 317 773 9082 USA and CANADA 1 800 743 5637 Cont...

Страница 2: ...r Cover I Error Codes Quality Control 18 A Calibration of the Temperature Controller B Alarm System Test Suggested Cleaning Procedure 20 Recommended Periodic Maintenance 21 Troubleshooting 22 Parts Li...

Страница 3: ...r is clean of any packaging debris and fill the chamber bath with distilled or deionized water until the water reaches the desired Fill Line There are two Fill Lines stamped in the center of the chamb...

Страница 4: ...the General Operation section Caution The technical complexities of the basket assembly lift out system require special handling and operational guidelines by the user Failure to follow the operationa...

Страница 5: ...____ C ____ 7 High alarm system check Signature________________________________Title____________________________ Department___________________________________________Date_______________ Controls and C...

Страница 6: ...rough each preprogrammed time setting If you wish to extend a thawing cycle that is in progress press this button the digital timer display will read HL and the thawing cycle will continue indefinitel...

Страница 7: ...een the two Time Set Buttons on the main control panel When the MUTE button is pressed the Cycle Time Indicators will flash on and off The High Alarm Indicator light will remain on until the alarm con...

Страница 8: ...the drain valve plug If you are draining the chamber water to the right side of the unit contact Helmer and request a U Fitting Assembly The U Fitting Assembly will allow for the drain tubing to run...

Страница 9: ...ycle Press the Start Stop Button to begin the thawing cycle The basket assembly will automatically lower into the chamber bath and begin agitating The Cycle Time Indicator s will begin to count down t...

Страница 10: ...1 360008 G 9 but will result in progressively slower thawing cycles if the bath water is not allowed to return to the set operating temperature...

Страница 11: ...1 360008 G 10...

Страница 12: ...alue 1 Operating Set Temp Press SET key Chamber Set Temperature 36 5 2 Operations Menu Press UP and DOWN keys together for 3 seconds Auto tune Low Alarm Deviation High Alarm Deviation Aut ALO Ah1 no O...

Страница 13: ...be ALO If the Aut prompt appears the controller programs are fully unlocked If the CnFg prompt appears the controller programs are fully lockout 2 Press the up or down arrow until the CnFg prompt appe...

Страница 14: ...t You are now at the program display for the Set Point Lockout 5 Press and hold the SET key to display the current input value of the Set Point program either YES or no will be displayed The YES promp...

Страница 15: ...1 360008 G 14...

Страница 16: ...m Programs To alter the Low and or High Alarm settings 1 Press and hold the Up and Down arrow keys until a controller program is displayed If the CnFg prompt appears the Controller Program Lockout wil...

Страница 17: ...y This will allow the drain tubing to be easily run to the right side of the unit G Plasma Overwraps Included with the unit is a dispenser box of 250 Helmer Plasma Overwraps The disposable Plasma Over...

Страница 18: ...respective problems Refer to the Troubleshooting section of the manual for further information on resolving problems that occur Control Panel Error Code Problem E1 The water temperature has exceeded t...

Страница 19: ...ow key until the PId Menu prompt appears 3 While pressing the SET key in the digital readout will display no press the Down arrow key to display yes and then release the SET key 4 Press the Up arrow o...

Страница 20: ...setting whereupon the visual and audible alarms should activate the basket assemblies should lift out of the chamber and the digital agitation timers should blink E1 4 After conducting this high alarm...

Страница 21: ...o be removed for cleaning when necessary 4 Using a soft cloth or sponge and a solution of distilled or deionized water with bath cleaner inhibitor thoroughly clean the chamber bath walls and basket as...

Страница 22: ...duty oil not grease on your finger and rubbing the oil up and down the rail Apply the thin coat of lubrication to each of the 4 sides on both lift out rails Troubleshooting Main Power Problem Possibl...

Страница 23: ...ater Verify voltage to heater Replace heater as needed Faulty circuit board Verify voltage to heater Replace circuit board as needed Faulty RTD sensor Verify sensor is adhered to chamber Verify connec...

Страница 24: ...rm not activating at high alarm setting Incorrect alarm function value Verify that the desired high alarm temperature value is correctly inputted into the controller Refer to Setting the High Alarm Ca...

Страница 25: ...ler wheel is loose Verify that the roller wheel is tight to the shaft of the lift out motor Tighten the roller wheel set screw Faulty connection Verify wiring connections from the circuit board to the...

Страница 26: ...Switches not in ON position Switch to ON position Faulty connections Verify that connections from the circuit board to the agitator motor assembly Loose cam wheel Verify that the cam wheel is securel...

Страница 27: ...assembly Verify that the finger screw on the top of the basket assembly is tight Basket assembly bearing Verify that the plastic roller bearings at the bottom of the basket assembly are operating pro...

Страница 28: ...t Button is pressed Push button not activating Verify Time Set Button activates when pressed replace as needed Verify that the circuit board is securely mounted to the chassis studs Faulty circuit boa...

Страница 29: ...itation motor 400257 115v 400258 230v 400257 100v Agitation motor gearbox 220231 Agitation motor capacitor 120047 115v 120301 230v 120047 100v Agitation cam wheel 1 320508 Agitation roller bearing blo...

Страница 30: ...1 360008 G 29...