Chapter 2
6
Convectron
®
ATM Vacuum Gauge Module Instruction Manual
Figure 2-2 illustrates the RS–485 version of the module. In the RS–485
version, each of the four setpoint relays can be independently assigned to
the Convectron gauge or the differential pressure sensor.
Figure 2-2
RS–485 version of module
2.2
Installation procedure
The module installation procedure includes the following steps:
1.
Installing appropriate pressure relief devices in the vacuum system.
2.
Locating and orienting the module.
3.
Attaching the module vacuum chamber fitting to its mate on the
vacuum chamber.
4.
Assembling and connecting module wiring.
5.
Configuring the setpoint relays to the desired voltage levels (analog
version) or digital pressure values (RS–485 version).
6.
Calibrating the Convectron gauge at atmospheric and vacuum
pressures.
This chapter also explains what to do if radio frequency interference (RFI)
disrupts operation of the RS–485 version of the module.
Vacuum chamber
fitting
Setpoint activation
indicators
Optional display
15-pin male
subminiature D
connector
WARNING
Failure to use accurate pressure conversion data for N
2
or air to other gases can cause an explosion due to
overpressurization.
If the module will measure any gas other than N
2
or air, before
putting the module into operation, adjust relays for the process
gas that will be used.
Содержание GRANVILLE-PHILLIPS Convectron ATM
Страница 2: ......
Страница 4: ......
Страница 8: ...iv Convectron ATM Vacuum Gauge Module Instruction Manual ...
Страница 12: ...4 Convectron ATM Vacuum Gauge Module Instruction Manual ...
Страница 80: ...72 Convectron ATM Vacuum Gauge Module Instruction Manual ...
Страница 90: ...82 Convectron ATM Vacuum Gauge Module Instruction Manual ...
Страница 97: ......