Testing the EPC9
EPC9 Manual
21
Put the model circuit into the ”10 MOhm” setting, which simulates a 10 M
Ω
-pipette
that is open to the bath solution. Reset the amplifier (1), set E9SCREEN to
VC
(”voltage clamp”) mode (2) and apply a test pulse of 5 mV amplitude (3) and 5 ms
duration (4). The current response will be displayed on the digital oscilloscope. If
your gain range is appropriate, i.e. 5 mV/pA (5) you should a see a rectangular
current of about 500 pA in response to the test pulse. This represents the ohmic
resistor you are recording from: I = U / R = 5 mV / 10 M = 500 pA. E9SCREEN
will online calculate the pipette resistance and update it in the
Rmem
field (6) where
you should read a value close to 10 M
Ω
.
A possible voltage offset can be automatically canceled by clicking on the
Auto-V0
button (7). After doing so, the command potential will be set to 0 mV (8) and the
V0
control (9) displays the offset potential. The baseline of the current response (10), the
voltage monitor
V-mon
(11) and the current monitor
I-mon
(12) should be close to
zero. You could also do the offset potential cancellation in a more classical way by
clicking into the
V0
control (9) and dragging the mouse up and down until the first
segment in the oscilloscope and the
I-mon
display (12) match zero.
The steps listed above can be automatically executed by clicking on the
SET-UP
button or pressing the
‘1’
key on the numerical keypad. This will execute the
Gentle CC-Switch: ON
Relative Value
Last V-memb
CC Fast Speed
Sound
Volume
Hz/M
Ω
Rmem
Sound
100 %
100
Scope
1. Amplifier
CC-range: 1pA/mV
Test Pulse
Amplitude
5.0 mV
Duration
5.0 ms
Imon2
Gain
1.02 x
Clear
Reset
Overlay
Close
Stimulus
External
Off
20 µs
Noise
0.00 A
Leak Subtraction
Off
Track
Auto
RsComp
Off
Off
CSlow
<>
1.00 pF
RSeries
5.0 M
Ω
Range
Off
Cap Track
Auto
CFast
<>
0.00 pF
50 %
0.5 µs
Auto
Filter 2
2.9 kHz
Bessel
Filter 1
Bessel 10 kHz
Rmem
10.0 M
Ω
V-mon
0.0 mV
I-mon
3.53 pA
Mode
VC
Vo
-2.8 mV
Track
Auto
LJ
0.0 mV
Gain
5.0 mV/pA
Clip
V-memb
0.0 mV
+/-
empty
empty
empty
empty
WHOLE-CELL
ON-CELL
SET-UP
Record
Delay
Off
Help
500. pA
1.20 ms
2
3
4
5
6
7
8
9
10
11
12
1
EPC9 Patch-Clamp Amplifier, v8.3