Operating Instructions 

)

)-

320 / 

)-

322

Process indicators for standard analog signals with 2 presets / relays and 2 control inputs 

 

Product features: 

x

Two identical models, but with a different power supply range

x

Analog input 0 … +/-10 V (or 2 – 10 V) and 0 – 20 mA (or 4 – 20 mA)

x

Two relay outputs for limit monitoring (dry changeover contacts)

x

Two digital multi-function inputs with programmable control function

x

Useful supplementary functions like Digital Filter, Linearization, Totalizer etc.

x

6 digit 14-segment display (14 mm / 0.55’’ size) for readout and menu dialogues

Available devices: 

x

)-

:  Power supply 10 – 30 VDC

x

)-

:

   

Power supply 90 – 240 VAC and auxiliary output 24 VDC

Tel.: +49 (0) 761 3841 3864
Fax.: +49 (0) 761 3841 3865

E-mail.: [email protected]
http://www.heinlanz.com

Hein Lanz GmbH

Elsaesserstrasse 15L, D-79110 
Freiburg im Breisgau, Germany

1

Содержание HL320

Страница 1: ...eover contacts x Two digital multi function inputs with programmable control function x Useful supplementary functions like Digital Filter Linearization Totalizer etc x 6 digit 14 segment display 14 mm 0 55 size for readout and menu dialogues Available devices x Power supply 10 30 VDC x Power supply 90 240 VAC and auxiliary output 24 VDC Tel 49 0 761 3841 3864 Fax 49 0 761 3841 3865 E mail info he...

Страница 2: ...Notes 4 2 Short Form Description 5 3 Display and Operating Elements 6 3 1 Overview 6 3 2 Special Key Functions 6 4 Block Diagram and Terminal Assignments 7 5 Operating Concept Mode of Operation 8 6 Programming 9 6 1 How to Enter the Menu Select a Function Group Quit the Menu 9 6 2 How to Select Functions Set Parameters Accept Settings 10 7 Function Groups 11 7 1 Help Texts running text 11 7 2 Sign...

Страница 3: ...r notice Therefore there might be minor differences between the unit and the descriptions in operation manual The raiser respectively positioner is exclusively responsible for the safety of the system and equipment where the unit will be integrated During installation operation or maintenance all general and also all country and application specific safety rules and standards must be observed If t...

Страница 4: ...the wires must be ensured Indications about the permissible wire cross sections for wiring are described in the Technical Specifications chapter Before first start up it must be ensured that all connections and wires are firmly seated and secured in the screw terminals All inclusively unused terminals must be fastened by turning the relevant screws clockwise up to the stop Overvoltages at the conn...

Страница 5: ...anguage for the Help Text selectable as English or German Signal input for 0 10 V 2 10 V 10 V 0 20 mA and 4 20 mA Sampling rate 10 readings per second Digital filter 1st order for smoothing display fluctuations with unstable input signals Customized linearization MIN MAX memory function Totalizer function 2 Relay outputs changeover contacts for limit monitoring Start delay for relay outputs after ...

Страница 6: ...us Function Group Select previous Function Decrement parameter value Select next Function Group Select next Function Increment parameter value Enter a Function Group Enter a Function Accept the new setting 4 S Space for unit overlay 3 2 Special Key Functions To enter the Programming Menu 3 sec To display Device Type and Software Version 5 sec To restore factory default settings 3 sec Tel 49 0 761 ...

Страница 7: ... Block Diagram and Terminal Assignments Tel 49 0 761 3841 3864 Fax 49 0 761 3841 3865 E mail info heinlanz com http www heinlanz com Hein Lanz GmbH Elsaesserstrasse 15L D 79110 Freiburg im Breisgau Germany ...

Страница 8: ...erating Concept Mode of Operation see 7 4 Tel 49 0 761 3841 3864 Fax 49 0 761 3841 3865 E mail info heinlanz com http www heinlanz com Hein Lanz GmbH Elsaesserstrasse 15L D 79110 Freiburg im Breisgau Germany ...

Страница 9: ...ng the programming menu via SAVE the minimum and maximum values and the totalizer value are all cleared 6 1 How to Enter the Menu Select a Function Group Quit the Menu Tel 49 0 761 3841 3864 Fax 49 0 761 3841 3865 E mail info heinlanz com http www heinlanz com Hein Lanz GmbH Elsaesserstrasse 15L D 79110 Freiburg im Breisgau Germany ...

Страница 10: ...t Parameters Accept Settings Fu unction Groups Func ction Parameter Tel 49 0 761 3841 3864 Fax 49 0 761 3841 3865 E mail info heinlanz com http www heinlanz com Hein Lanz GmbH Elsaesserstrasse 15L D 79110 Freiburg im Breisgau Germany ...

Страница 11: ...nge 10 10 V Range 0 20 mA Range 4 20 mA Select l lower measuring range limit Setting range depends on measuring range V mA Select upper measuring range limit Setting range depends on measuring range V mA Select Decimal Point visual function only 0 0 0 0 00 0 000 0 0000 Select input low value Setting range depends on measuring range V mA Select low display value for I INP LO Setting range 199999 99...

Страница 12: ...ge 0 10 V 0 500 10 500 V 2 10 V 1 500 10 500 V 10 10 V 10 500 10 500 V 0 20 mA 0 500 21 000 mA 4 20 mA 3 000 21 000 mA x The functions LO LIM and HI LIM will limit the editable range correspondingly x The setting of H HI LIM has to be greater than the setting of L LO LIM a at any time 7 3 User Linearization Menu u U User L Linearization Select User Linearization Linearization OFF Linearization ONF...

Страница 13: ...alue from the analog input in physical units as well as the corresponding display value DISP for this sensor value x The linearization points LP may be entered in any order They are then sorted in the firmware in ascending order for the linearization function x Linearization based on a negative slope is possible x The setting range depends on the selected measuring range 0 10 V 0 500 10 500 V 2 10...

Страница 14: ... Example 2 4 20 mA 300 900 E Example3 1 0 9 0 V 750 400 7 3 2 Example for non linear scaling Tel 49 0 761 3841 3864 Fax 49 0 761 3841 3865 E mail info heinlanz com http www heinlanz com Hein Lanz GmbH Elsaesserstrasse 15L D 79110 Freiburg im Breisgau Germany 1 ...

Страница 15: ...ut 2 Reset of relay outputs in Latch mode x In the function group MP KEY programme the function RES REL to ON In the operating mode briefly press the MP Key x In the function group MP INP programme the function MP INP1 or MP INP2 to RES REL In the operating mode briefly activate the multifunction input 1 or multifunction input 2 Reset Totalizer x In the function group MP KEY programme the function...

Страница 16: ...y when MAX value is in display OFF ON Select function Output Latch Reset by MP key only if output is in memory mode ALARMx LATCH OFF ON this setting is possible only if all other settings are set to OFF Select function Reset Totalizer by MP key when the totalizer value is in display OFF ON Tel 49 0 761 3841 3864 Fax 49 0 761 3841 3865 E mail info heinlanz com http www heinlanz com Hein Lanz GmbH E...

Страница 17: ... Key Tare Reset tare value Reset totalizer Select fu unction MP input 2 No function Reset MIN value Reset MAX value Reset MIN and MAX values Reset Output Latch only if output is in memory mode ALARMx LATCH Hold freeze display Lock out limit value setting Lock out programming and default setting Lock out limit value setting programming default setting and MP Key Tare Reset tare value Reset totalize...

Страница 18: ...ery 100 msec From these values the throughput rate per second is calculated To convert the throughput rate into other units FACTOR and SCALE are available Negative measured values are also taken into consideration A decimal point that is programmed with a current measured value is not taken into consideration The throughput rate is added into the totalizer memory every second A dummy decimal point...

Страница 19: ... Tel 49 0 761 3841 3864 Fax 49 0 761 3841 3865 E mail info heinlanz com http www heinlanz com Hein Lanz GmbH Elsaesserstrasse 15L D 79110 Freiburg im Breisgau Germany 1 ...

Страница 20: ...put active With alarm output inactive Select O ON h hysteresis Setting range 0 9999 and decimal point Select O OFF h hysteresis o only with auto operation Setting range 0 9999 and decimal point Select O ON d delay Setting range 0 0 99 9 sec Select O OFF d delay only with auto operation Input range 0 0 99 9 sec Menu Alarm Output 2 2 Select operating mode OFF Automatic operation Memory latch operati...

Страница 21: ...hysteresis OFF switching point limit value OFF hysteresis An alarm is triggered if the measured value falls outside a defined range Band Upper switching point limit value ON hysteresis Lower switching point limit value OFF hysteresis An alarm causes the output to become active Relay energized LED ON An alarm causes the output to become inactive Relay not energized LED OFF In memory latch operation...

Страница 22: ...easuring signal 7 7 6 3 C Control with d dec crementing measuring signal Tel 49 0 761 3841 3864 Fax 49 0 761 3841 3865 E mail info heinlanz com http www heinlanz com Hein Lanz GmbH Elsaesserstrasse 15L D 79110 Freiburg im Breisgau Germany ...

Страница 23: ...E 8E A 9 E B59 D 5 5 EA 9 E BE A 9 9E B59 2 5 5E EA 9 9E BE 9 B A 9 E B59 D 5 5 EA 9 E BE 9 9 E BE D 9 9 E 0 9E Tel 49 0 761 3841 3864 Fax 49 0 761 3841 3865 E mail info heinlanz com http www heinlanz com Hein Lanz GmbH Elsaesserstrasse 15L D 79110 Freiburg im Breisgau Germany ...

Страница 24: ...nge Limit Upper Meas R Range Limit Probe or wire s short circuit Probe or wire b break 0 10 V 2 10 V 1 V 10 V 10 V 0 20 mA 4 20 mA 2 mA Indication blinking blinking blinking blinking blinking is detected Tel 49 0 761 3841 3864 Fax 49 0 761 3841 3865 E mail info heinlanz com http www heinlanz com Hein Lanz GmbH Elsaesserstrasse 15L D 79110 Freiburg im Breisgau Germany ...

Страница 25: ... max 0 05 v Mb Temperature drift 100 ppm K Input resistance 22 Ω PTC 25 Ω Voltage drop ca 1 8 V 20 mA Maximum current max 60 mA Control inputs M MPI 1 MPI 2 Quantity 2 optocoupler inputs with programmable function Switching levels Low 2 V High 4 V max 30 V Pulse length 100 ms Alarm outputs Relays dry change over contacts Switching voltage max 250 V AC 125 V DC min 5 V AC V DC Switching current max...

Страница 26: ...ion level 2 M Mechanical data Housing Panel mount housing to DIN 43 700 RAL 7021 Dimensions 96 x 48 x 102 mm 3 78 x 1 89 x 4 02 Panel cut out 92 0 8 x 45 0 6 mm 3 62 0 03 x 1 77 0 23 Installation depth ca 92 mm 3 62 including terminals Weight ca 180 g Protection IP 65 front Housing material Polycarbonate UL94 V 2 Vibration resistance EN60068 2 6 10 55 Hz 1 mm XYZ 30 min in each direction Shock res...

Страница 27: ... Panel Cut Out 92 0 8 x x 45 0 6 m mm 3 62 0 03 x 1 77 0 23 Tel 49 0 761 3841 3864 Fax 49 0 761 3841 3865 E mail info heinlanz com http www heinlanz com Hein Lanz GmbH Elsaesserstrasse 15L D 79110 Freiburg im Breisgau Germany ...

Страница 28: ...UE INP HI INPUT END VALUE DISP HI DISPLAY END VALUE FILTER INPUT FILTER PW DELY POWER ON DELAY FOR OUTPUTS SEC PW FREQ 50HZ POWER LINE FREQUENCY 50HZ PW FREQ 60HZ POWER LINE FREQUENCY 60HZ LINEAR MAIN MENU LINEARIZATION LINEAR NO LINEARIZATION OFF LINEAR YES LINEARIZATION ON NUM PNT NUMBER OF LINEARIZATION POINTS INP 01 INPUT VALUE NO 1 DISP 01 DISPLAY VALUE NO 1 to INP 10 INPUT VALUE NO 10 DISP 1...

Страница 29: ...OR TOTALIZER X0 0001 DP TOT 0 NO DECIMAL POINT DP TOT 0 0 DECIMAL POINT TOTALISER 0 0 DP TOT 0 00 DECIMAL POINT TOTALISER 0 00 DP TOT 0 000 DECIMAL POINT TOTALISER 0 000 DP TOT 0 0000 DECIMAL POINT TOTALISER 0 0000 DP TOT 0 00000 DECIMAL POINT TOTALISER 0 00000 ALARMx MAIN MENU ALARM x AL OUTx OFF ALARM x OFF AL OUTx AUTO AUTOMATIC MODE OF ALARM OUTPUT x AL OUTx LATCH LATCH MODE OF ALARM OUTPUT x ...

Страница 30: ...ERFLOW OVERFL OVERFLOW UPPER INPUT RANGE LIMIT UNDERF UNDERFLOW LOWER INPUT RANGE LIMIT Tel 49 0 761 3841 3864 Fax 49 0 761 3841 3865 E mail info heinlanz com http www heinlanz com Hein Lanz GmbH Elsaesserstrasse 15L D 79110 Freiburg im Breisgau Germany ...